Clinical correlation between a point-of-care testing system and laboratory automation for lipid profile

Clin Chim Acta. 2015 Jun 15:446:263-6. doi: 10.1016/j.cca.2015.04.036. Epub 2015 May 4.

Abstract

Background: We evaluated the clinical correlation between the CardioChek PA analyzer and a clinical laboratory reference method to use for screening program purposes.

Methods: Fasting blood samples were collected on 516 patients (age 20-85 y). One venous sample was collected using a serum tube for the evaluation on a COBAS reference analyzer. A second venous sample was collected in a lithium heparin tube and was evaluated on the CardioChek PA analyzer (CCPA venous). A fingerstick sample (CCPA fingerstick) was evaluated only on the CardioChek PA analyzer. Linear regression analyses were performed for each measured analyte, total cholesterol, HDL-cholesterol and triglycerides.

Results: The correlation between the CCPA fingerstick and CCPA venous was extremely high for HDL-C and triglycerides, and good for total cholesterol. Our results demonstrated a good clinical agreement for total cholesterol, HDL-C and triglycerides between 97.7% and 94.6% in the comparison of the CCPA to the reference analyzer.

Conclusions: We identified the pre-analytic phase as an important step to guarantee the quality of results and indicate that the CardioChek PA is a reliable lipid point-of-care testing system that can be used for the application of clinical screening anywhere.

Keywords: CardioChek PA; Cardiovascular diseases; Point-of-care testing.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Automation, Laboratory / standards*
  • Cardiovascular Diseases / blood*
  • Cardiovascular Diseases / diagnosis
  • Cholesterol, HDL / blood*
  • Fasting / blood
  • Female
  • Humans
  • Linear Models
  • Male
  • Middle Aged
  • Point-of-Care Testing / standards*
  • Practice Guidelines as Topic
  • Triglycerides / blood*

Substances

  • Cholesterol, HDL
  • Triglycerides