Objective: This study aimed to evaluate the accuracy of the Andon KD-5965 upper-arm blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010.
Methodology: Systolic and diastolic blood pressures were sequentially measured in 33 adults, with 20 women using a mercury sphygmomanometer (two observers) and the Andon KD-5965 device (one supervisor). A total of 99 pairs of comparisons were obtained from 33 participants for judgments in two parts with three grading phases.
Results: The device achieved the targets in part 1 of the validation study. The number of absolute differences between the device and observers within 5, 10, and 15 mmHg was 70/99, 91/99, and 98/99, respectively, for systolic blood pressure and 81/99, 99/99, and 99/99, respectively, for diastolic blood pressure. The device also fulfilled the criteria in part 2 of the validation study. Twenty-five and 29 participants, for systolic and diastolic blood pressure, respectively, had at least two of the three device-observers differences within 5 mmHg (required≥24). Two and one participants for systolic and diastolic blood pressure, respectively, had all three device-observers comparisons greater than 5 mmHg.
Conclusion: According to the validation results, with better performance for diastolic blood pressure than that for systolic blood pressure, the Andon automated oscillometric upper-arm blood pressure monitor KD-5965 fulfilled the requirements of the European Society of Hypertension International Protocol revision 2010, and hence can be recommended for blood pressure measurement in adults.