A Novel Configuration of Pipeline Embolization Device for Internal Carotid Bifurcation Region Aneurysms: Horizontal Deployment

Clin Neuroradiol. 2017 Mar;27(1):57-60. doi: 10.1007/s00062-015-0414-y. Epub 2015 Jun 6.

Abstract

Purpose: To describe a novel configuration of pipeline embolization device for internal carotid bifurcation region aneurysm, named horizontal stenting.

Clinical presentation: A 64-year-old woman, with visual deficit, harboring a large wide-necked aneurysm located at the junction between left internal carotid artery and left A1 segment of anterior cerebral artery, was submitted to endovascular treatment. As she had pre-existing occlusion of left internal carotid, approach from the contralateral internal carotid was used to advance the pipeline embolization device through the anterior communicating artery and place the flow diverter horizontally across the neck (from M1 to A1). Coil embolization was also performed through a microcatheter navigated via posterior communicating artery. The intervention was uneventful, with total aneurysm occlusion. Patient presented with visual improvement on follow-up.

Conclusion: Horizontal deployment of pipeline embolization device appears to be an acceptable and feasible alternative to treat internal carotid bifurcation aneurysms. Long-term follow-up and a greater number of cases are mandatory to establish the safety of this strategy.

Keywords: Aneurysm; Embolization; Stenting.

Publication types

  • Case Reports

MeSH terms

  • Carotid Artery Diseases / diagnostic imaging
  • Carotid Artery Diseases / surgery*
  • Embolization, Therapeutic / instrumentation*
  • Embolization, Therapeutic / methods*
  • Equipment Failure Analysis
  • Female
  • Humans
  • Intracranial Aneurysm / diagnostic imaging
  • Intracranial Aneurysm / surgery*
  • Middle Aged
  • Prosthesis Design
  • Prosthesis Implantation / methods*
  • Treatment Outcome