Description and comparison of algorithms for correcting anisotropic magnification in cryo-EM images
- PMID: 26087140
- DOI: 10.1016/j.jsb.2015.06.014
Description and comparison of algorithms for correcting anisotropic magnification in cryo-EM images
Abstract
Single particle electron cryomicroscopy (cryo-EM) allows for structures of proteins and protein complexes to be determined from images of non-crystalline specimens. Cryo-EM data analysis requires electron microscope images of randomly oriented ice-embedded protein particles to be rotated and translated to allow for coherent averaging when calculating three-dimensional (3D) structures. Rotation of 2D images is usually done with the assumption that the magnification of the electron microscope is the same in all directions. However, due to electron optical aberrations, this condition is not met with some electron microscopes when used with the settings necessary for cryo-EM with a direct detector device (DDD) camera. Correction of images by linear interpolation in real space has allowed high-resolution structures to be calculated from cryo-EM images for symmetric particles. Here we describe and compare a simple real space method, a simple Fourier space method, and a somewhat more sophisticated Fourier space method to correct images for a measured anisotropy in magnification. Further, anisotropic magnification causes contrast transfer function (CTF) parameters estimated from image power spectra to have an apparent systematic astigmatism. To address this problem we develop an approach to adjust CTF parameters measured from distorted images so that they can be used with corrected images. The effect of anisotropic magnification on CTF parameters provides a simple way of detecting magnification anisotropy in cryo-EM datasets.
Keywords: Anisotropic magnification; Contrast transfer function; Correction; Cryo-EM; Differential magnification; Direct detector device; Direct electron detector; Distortion.
Copyright © 2015 Elsevier Inc. All rights reserved.
Similar articles
-
An algorithm for estimation and correction of anisotropic magnification distortion of cryo-EM images without need of pre-calibration.J Struct Biol. 2016 Aug;195(2):207-215. doi: 10.1016/j.jsb.2016.06.003. Epub 2016 Jun 4. J Struct Biol. 2016. PMID: 27270241 Free PMC article.
-
Automatic estimation and correction of anisotropic magnification distortion in electron microscopes.J Struct Biol. 2015 Nov;192(2):204-8. doi: 10.1016/j.jsb.2015.08.006. Epub 2015 Aug 13. J Struct Biol. 2015. PMID: 26278979 Free PMC article.
-
Alignment algorithms and per-particle CTF correction for single particle cryo-electron tomography.J Struct Biol. 2016 Jun;194(3):383-94. doi: 10.1016/j.jsb.2016.03.018. Epub 2016 Mar 22. J Struct Biol. 2016. PMID: 27016284 Free PMC article.
-
Processing of Cryo-EM Movie Data.Methods Enzymol. 2016;579:103-24. doi: 10.1016/bs.mie.2016.04.009. Epub 2016 Jun 1. Methods Enzymol. 2016. PMID: 27572725 Review.
-
Cryo-electron microscopy and the amazing race to atomic resolution.Biochemistry. 2015 May 26;54(20):3133-41. doi: 10.1021/acs.biochem.5b00114. Epub 2015 May 14. Biochemistry. 2015. PMID: 25955078 Review.
Cited by
-
Models for the a subunits of the Thermus thermophilus V/A-ATPase and Saccharomyces cerevisiae V-ATPase enzymes by cryo-EM and evolutionary covariance.Proc Natl Acad Sci U S A. 2016 Mar 22;113(12):3245-50. doi: 10.1073/pnas.1521990113. Epub 2016 Mar 7. Proc Natl Acad Sci U S A. 2016. PMID: 26951669 Free PMC article.
-
Structure and conformational states of the bovine mitochondrial ATP synthase by cryo-EM.Elife. 2015 Oct 6;4:e10180. doi: 10.7554/eLife.10180. Elife. 2015. PMID: 26439008 Free PMC article.
-
Structure of a AAA+ unfoldase in the process of unfolding substrate.Elife. 2017 Apr 8;6:e25754. doi: 10.7554/eLife.25754. Elife. 2017. PMID: 28390173 Free PMC article.
-
Electron cryomicroscopy observation of rotational states in a eukaryotic V-ATPase.Nature. 2015 May 14;521(7551):241-5. doi: 10.1038/nature14365. Nature. 2015. PMID: 25971514
-
Molecular basis for the binding and modulation of V-ATPase by a bacterial effector protein.PLoS Pathog. 2017 Jun 1;13(6):e1006394. doi: 10.1371/journal.ppat.1006394. eCollection 2017 Jun. PLoS Pathog. 2017. PMID: 28570695 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous