Stability of eosin-5'-maleimide dye used in flow cytometric analysis for red cell membrane disorders

Blood Res. 2015 Jun;50(2):109-12. doi: 10.5045/br.2015.50.2.109. Epub 2015 Jun 25.

Abstract

Background: The eosin-5'-maleimide (EMA) binding test using flow cytometry is a common method to measure reduced mean channel fluorescence (MCF) of EMA-labeled red blood cells (RBCs) from patients with red cell membrane disorders. The basic principle of the EMA-RBC binding test involves the covalent binding of EMA to lysine-430 on the first extracellular loop of band 3 protein.

Methods: In the present study, the MCF of EMA was analyzed for samples derived from 12 healthy volunteers (controls) to determine the stability (i.e., the percentage decrease in fluorescence) of EMA over a period of 1 year.

Results: Comparison of periodical MCF readings over time, that is, at 2-month intervals, showed that there were no significant changes in mean channel fluorescence for up to 6 months; however, there was a significant decrease in MCF at 8 months.

Conclusion: For optimal dye utilization, EMA remained stable only for up to 6 months. Therefore, we recommend reconstitution of the dye every 6 months when implementing this test and storage at -80℃ in dark conditions.

Keywords: Flow Cytometry; Hereditary spherocytosis; Red Blood cells.