Comparison of Anterior Segment Measurements With a Dual Scheimpflug Placido Corneal Topographer and a New Partial Coherence Interferometer in Keratoconic Eyes

Cornea. 2015 Sep;34(9):1012-8. doi: 10.1097/ICO.0000000000000480.

Abstract

Purpose: To evaluate the repeatability and agreement of the anterior segment measurements obtained using the Galilei dual Scheimpflug analyzer (Galilei DSA; Ziemer) and Nidek AL Scan (Nidek CO, Aichi, Japan) biometry in keratoconic and normal eyes.

Methods: Three consecutive measurements were performed by the same examiner using both devices in 62 healthy and 88 keratoconic eyes. Central corneal thickness (CCT), anterior chamber depth, keratometry readings (K) [in flattest meridian (Kf), in steepest meridian (Ks), and mean (Km)], and white-to-white distance were evaluated. Repeatability was assessed by calculating the within-subject SD and coefficient of variation. The agreement between both devices was assessed using the Bland-Altman method.

Results: Both devices achieved excellent repeatability for all parameters in each group. The 95% limits of agreement (LoA) between both devices were also very narrow and acceptable for all parameters in normal corneas. However, the 95% LoA for agreement was large for CCT and measurements related to K (Kf, Ks, and Km) using both 2.4 and 3.3 mm in keratoconic eyes. In addition, compared with the Galilei DSA, K values of the Nidek AL Scan using a diameter of 3.3 mm showed slightly closer 95% LoA than those obtained using a diameter of 2.4 mm.

Conclusions: In normal eyes, the Galilei DSA and Nidek AL Scan can be used interchangeably for anterior segment measurements. In keratoconic eyes, both devices yielded interchangeable anterior chamber depth and white-to-white distance measurements, whereas CCT and keratometry measurements showed clinically significant differences.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Anterior Eye Segment / anatomy & histology
  • Anterior Eye Segment / pathology*
  • Biometry
  • Corneal Topography / instrumentation*
  • Female
  • Healthy Volunteers
  • Humans
  • Interferometry / instrumentation*
  • Keratoconus / classification
  • Keratoconus / diagnosis*
  • Light
  • Male
  • Prospective Studies
  • Reproducibility of Results
  • Young Adult