Insulation lifetime improvement of polyimide thin film neural implants

J Neural Eng. 2015 Oct;12(5):054001. doi: 10.1088/1741-2560/12/5/054001. Epub 2015 Aug 13.

Abstract

Objective: This work deals with studying and improving the insulation lifetime of polyimide-insulated thin film neural implants, or related polyimide-based medical implants.

Approach: The evolution of the leak impedance of insulated conductors was investigated in saline water at 40 °C. The fabrication process as commonly found in literature for polyimide/platinum/polyimide microelectrode arrays was compared with three possible improvements: one based on lowering the curing temperature of the lower layer, one based on chemical activation and one based on an additional plasma activation step.

Main results: The lower curing temperature process was found to yield a 7.5-fold improved lifetime compared with the state of the art process. Also, the leak impedances found after soak testing are an order of magnitude higher compared to the standard process.

Significance: By improving the lifetime and insulation impedance of polyimide insulation with one order of magnitude, this work increases the applicability of polyimide in chronic thin film neural implants considerably.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Body Water / chemistry*
  • Electric Impedance
  • Electrodes, Implanted*
  • Equipment Design
  • Equipment Failure Analysis
  • Materials Testing
  • Microelectrodes*
  • Neural Prostheses*
  • Resins, Synthetic / chemistry*

Substances

  • Resins, Synthetic
  • polyimide resin