1-Year Outcomes After Transcatheter Aortic Valve Replacement With Balloon-Expandable Versus Self-Expandable Valves: Results From the CHOICE Randomized Clinical Trial

J Am Coll Cardiol. 2015 Aug 18;66(7):791-800. doi: 10.1016/j.jacc.2015.06.026.

Abstract

Background: The use of a balloon-expandable transcatheter heart valve previously resulted in a greater rate of device success compared with a self-expandable transcatheter heart valve.

Objectives: The aim of this study was to evaluate clinical and echocardiographic outcome data at longer term follow-up.

Methods: The investigator-initiated trial randomized 241 high-risk patients with symptomatic severe aortic stenosis and anatomy suitable for treatment with both balloon- and self-expandable transcatheter heart valves to transfemoral transcatheter aortic valve replacement with either device. Patients were followed-up for 1 year, with assessment of clinical outcomes and echocardiographic evaluation of valve function.

Results: At 1 year, the rates of death of any cause (17.4% vs. 12.8%; relative risk [RR]: 1.35; 95% confidence interval [CI]: 0.73 to 2.50; p = 0.37) and of cardiovascular causes (12.4% vs. 9.4%; RR: 1.32; 95% CI: 0.63 to 2.75; p = 0.54) were not statistically significantly different in the balloon- and self-expandable groups, respectively. The frequencies of all strokes (9.1% vs. 3.4%; RR: 2.66; 95% CI: 0.87 to 8.12; p = 0.11) and repeat hospitalization for heart failure (7.4% vs. 12.8%; RR: 0.58; 95% CI: 0.26 to 1.27; p = 0.19) did not statistically significantly differ between the 2 groups. Elevated transvalvular gradients during follow-up were observed in 4 patients in the balloon-expandable group (3.4% vs. 0%; p = 0.12); all were resolved with anticoagulant therapy, suggesting a thrombotic etiology. More than mild paravalvular regurgitation was more frequent in the self-expandable group (1.1% vs. 12.1%; p = 0.005).

Conclusions: Despite the higher device success rate with the balloon-expandable valve, 1-year follow-up of patients in CHOICE (Randomized Comparison of Transcatheter Heart Valves in High Risk Patients With Severe Aortic Stenosis: Medtronic CoreValve Versus Edwards SAPIEN XT Trial), with limited statistical power, revealed clinical outcomes after transfemoral transcatheter aortic valve replacement with both balloon- and self-expandable prostheses that were not statistically significantly different. (A Comparison of Transcatheter Heart Valves in High Risk Patients With Severe Aortic Stenosis: The CHOICE Trial; NCT01645202).

Keywords: TAVR; aortic stenosis; balloon-expandable; self-expandable.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve Stenosis / diagnosis*
  • Aortic Valve Stenosis / surgery*
  • Balloon Valvuloplasty / standards
  • Balloon Valvuloplasty / trends*
  • Female
  • Follow-Up Studies
  • Heart Valve Prosthesis / standards
  • Heart Valve Prosthesis / trends*
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Prosthesis Design / standards
  • Prosthesis Design / trends*
  • Time Factors
  • Transcatheter Aortic Valve Replacement / standards
  • Transcatheter Aortic Valve Replacement / trends*
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT01645202