The Contribution of Common Surgically Implanted Hardware to Functional MR Imaging Artifacts

AJNR Am J Neuroradiol. 2015 Nov;36(11):2068-73. doi: 10.3174/ajnr.A4419. Epub 2015 Aug 13.

Abstract

Background and purpose: Blood oxygenation level-dependent MR imaging is increasingly used clinically to noninvasively assess cerebrovascular reactivity and/or language and motor function. However, many patients have metallic implants, which will induce susceptibility artifacts, rendering the functional information uninformative. Here, we calculate and interpret blood oxygenation level-dependent MR imaging artifact impact arising from surgically implanted hardware.

Materials and methods: A retrospective analysis of all blood oxygenation level-dependent MRIs (n = 343; B0 = 3T; TE = 35 ms; gradient echo EPI) acquired clinically (year range = 2006-2014) at our hospital was performed. Blood oxygenation level-dependent MRIs were most commonly prescribed for patients with cerebrovascular disease (n = 80) or patients undergoing language or motor localization (n = 263). Artifact volume (cubic centimeters) and its impact on clinical interpretation were determined by a board-certified neuroradiologist.

Results: Mean artifact volume associated with intracranial hardware was 4.3 ± 3.2 cm(3) (range = 1.1-9.4 cm(3)). The mean artifact volume from extracranial hardware in patients with cerebrovascular disease was 28.4 ± 14.0 cm(3) (range = 6.1-61.7 cm(3)), and in patients with noncerebrovascular disease undergoing visual or motor functional mapping, it was 39.9 (3)± 27.0 cm(3) (range = 6.9-77.1 cm(3)). The mean artifact volume for ventriculoperitoneal shunts was 95.7 ± 39.3 cm(3) (range = 64.0-139.6 cm(3)). Artifacts had no-to-mild effects on clinical interpretability in all patients with intracranial implants. Extracranial hardware artifacts had no-to-moderate impact on clinical interpretability, with the exception of 1 patient with 12 KLS-Martin maxDrive screws with severe artifacts precluding clinical interpretation. All examined ventriculoperitoneal shunts resulted in moderate-to-severe artifacts, limiting clinical interpretation.

Conclusions: Blood oxygenation level-dependent MR imaging yields interpretable functional maps in most patients beyond a small (30-40 cm(3)) artifact surrounding the hardware. Exceptions were ventriculoperitoneal shunts, particularly those with programmable valves and siphon gauges, and large numbers of KLS-Martin maxDrive screws.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Artifacts*
  • Brain / pathology*
  • Female
  • Functional Neuroimaging / methods*
  • Humans
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Prostheses and Implants*
  • Retrospective Studies