Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Multicenter Study
. 2015 Oct;242(2):433-42.
doi: 10.1016/j.atherosclerosis.2015.08.001. Epub 2015 Aug 5.

Reliability of measurement of endothelial function across multiple institutions and establishment of reference values in Japanese

Affiliations
Free article
Multicenter Study

Reliability of measurement of endothelial function across multiple institutions and establishment of reference values in Japanese

Hirofumi Tomiyama et al. Atherosclerosis. 2015 Oct.
Free article

Abstract

Aims: For the standardization of flow-mediated vasodilatation (FMD) assessment as a clinical tool, validation of its reliability across multiple institutions and the establishment of normal/reference values based on reliable data from multiple institutions are needed.

Methods and results: In Study 1, assessment of FMD (scan recording and analysis) using an ultrasonographic semi-automatic measuring system (sFMD) was conducted at 18 participating institutions (sFMD-INST) (n = 981). All of the brachial arterial scans were also analyzed at a core laboratory (sFMD-COLB). After 111 subjects with inadequate sFMD recordings were excluded (n = 880), the correlation between the sFMD-INST and sFMD-COLB improved from R = 0.725 to R = 0.838 (p < 0.001). In Study 2, based on good-quality sFMD data obtained from 6660 subjects without cardiovascular disease (CVD) and 729 subjects with CVD from 27 institutions, reference values of sFMD are proposed by the Framingham risk score (FRS)-based risk categories and according to gender and age. The receiver-operating characteristic curve analysis revealed a significant power of sFMD values in reference ranges to discriminate between subjects with and without CVD (e.g., area under curve = 0.64 in the FRS-low risk group).

Conclusions: When the analysis was limited to cases with clear sFMD recordings, the reliability of the sFMD assessment (scan and its analysis) conducted in individual institutions appeared to be acceptable. Reference sFMD values (lower cuff occlusion) for the Japanese population are proposed based on reliable data derived from multiple institutions, and the reference values may identify patients without advanced vascular damage.

Keywords: Endothelial function; Reference value; Reliability; Standardization.