Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF-1000i

J Clin Lab Anal. 2016 Sep;30(5):381-91. doi: 10.1002/jcla.21866. Epub 2015 Aug 24.

Abstract

Background: We evaluated the new body fluid module on Sysmex UF1000-i (UF1000i-BF) for analysis of white blood cell (WBC) and red blood cell (RBC) in cerebrospinal fluid (CSF).

Methods: WBC and RBC counting were compared between UF1000i-BF and Fuchs-Rosenthal counting chamber in 67 CSF samples. This study also included the evaluation of between-day precision, limit of blank (LoB), limit of detection (LoD), functional sensitivity (limit of quantitation, LoQ), carryover and linearity. Diagnostic agreement for differentiation between normal and increased WBC counts (≥5.0 × 10(6) /L) was also assessed.

Results: The agreement between UF1000i-BF and manual WBC counts was otpiaml in all CSF samples (r = 0.99; y = 1.05x + 0.09). A modest overestimation was noticed in samples with WBC < 30 × 10(6) /L (r = 0.95; y = 1.21x - 0.15). A good agreement was observed for RBC counts (r = 0.98; y = 1.15x + 0.55), particularly in samples with RBC ≥ 18 × 10(6) /L (r = 0.98; y = 1.01x + 8.90). Between-day precision was good, with coefficient of variations (CVs) lower than 7.2% for both WBC and RBC. The LoBs were 0.1 × 10(6) WBC/L and 1.2 × 10(6) RBC/L, the LoDs were 0.7 × 10(6) WBC/L and 5.5 × 10(6) RBC/L, the LoQs were 2.4 × 10(6) WBC/L and 18.0 × 10(6) RBC/L, respectively. Linearity was excellent (r = 1.00 for both WBC and RBC). Carryover was negligible. Excellent diagnostic agreement was obtained at 4.5 × 10(6) WBC/L cut-off (sensitivity, 100%; specificity, 97.4%).

Conclusion: The UF1000i-BF provides rapid and accurate WBC and RBC counts in clinically relevant values of CSF cells. The use of UF1000i-BF may hence allow to replace routine optical counting, except for samples displaying abnormal WBC counts or abnormal scattergram distribution, for which differential cell counts may still be required.

Keywords: automated cell counting; body fluids Sysmex UF-1000i-BF; cerebrospinal fluid; red blood cells; white blood cells.

MeSH terms

  • Adolescent
  • Adult
  • Area Under Curve
  • Automation, Laboratory / instrumentation
  • Automation, Laboratory / methods*
  • Body Fluids / cytology*
  • Cerebrospinal Fluid / cytology*
  • Erythrocyte Count*
  • Humans
  • Leukocyte Count
  • Leukocytes*
  • Linear Models
  • Male
  • Paracentesis
  • Young Adult