Percutaneous Implantation of an Entirely Intracardiac Leadless Pacemaker
- PMID: 26321198
- DOI: 10.1056/NEJMoa1507192
Percutaneous Implantation of an Entirely Intracardiac Leadless Pacemaker
Abstract
Background: Cardiac pacemakers are limited by device-related complications, notably infection and problems related to pacemaker leads. We studied a miniaturized, fully self-contained leadless pacemaker that is nonsurgically implanted in the right ventricle with the use of a catheter.
Methods: In this multicenter study, we implanted an active-fixation leadless cardiac pacemaker in patients who required permanent single-chamber ventricular pacing. The primary efficacy end point was both an acceptable pacing threshold (≤2.0 V at 0.4 msec) and an acceptable sensing amplitude (R wave ≥5.0 mV, or a value equal to or greater than the value at implantation) through 6 months. The primary safety end point was freedom from device-related serious adverse events through 6 months. In this ongoing study, the prespecified analysis of the primary end points was performed on data from the first 300 patients who completed 6 months of follow-up (primary cohort). The rates of the efficacy end point and safety end point were compared with performance goals (based on historical data) of 85% and 86%, respectively. Additional outcomes were assessed in all 526 patients who were enrolled as of June 2015 (the total cohort).
Results: The leadless pacemaker was successfully implanted in 504 of the 526 patients in the total cohort (95.8%). The intention-to-treat primary efficacy end point was met in 270 of the 300 patients in the primary cohort (90.0%; 95% confidence interval [CI], 86.0 to 93.2, P=0.007), and the primary safety end point was met in 280 of the 300 patients (93.3%; 95% CI, 89.9 to 95.9; P<0.001). At 6 months, device-related serious adverse events were observed in 6.7% of the patients; events included device dislodgement with percutaneous retrieval (in 1.7%), cardiac perforation (in 1.3%), and pacing-threshold elevation requiring percutaneous retrieval and device replacement (in 1.3%).
Conclusions: The leadless cardiac pacemaker met prespecified pacing and sensing requirements in the large majority of patients. Device-related serious adverse events occurred in approximately 1 in 15 patients. (Funded by St. Jude Medical; LEADLESS II ClinicalTrials.gov number, NCT02030418.).
Comment in
-
Device therapy: Leadless pacemaker demonstrates safety, efficacy, and retrievability.Nat Rev Cardiol. 2015 Nov;12(11):620. doi: 10.1038/nrcardio.2015.140. Epub 2015 Sep 15. Nat Rev Cardiol. 2015. PMID: 26370476 No abstract available.
-
Achilles' Lead: Will Pacemakers Break Free?N Engl J Med. 2016 Feb 11;374(6):585-6. doi: 10.1056/NEJMe1513625. Epub 2015 Nov 9. N Engl J Med. 2016. PMID: 26551666 No abstract available.
-
A Leadless Cardiac Pacemaker.N Engl J Med. 2016 Feb 11;374(6):594. doi: 10.1056/NEJMc1515015. N Engl J Med. 2016. PMID: 26863365 No abstract available.
-
A Leadless Cardiac Pacemaker.N Engl J Med. 2016 Feb 11;374(6):593. doi: 10.1056/NEJMc1515015. N Engl J Med. 2016. PMID: 26863366 No abstract available.
-
A Leadless Cardiac Pacemaker.N Engl J Med. 2016 Feb 11;374(6):593-4. doi: 10.1056/NEJMc1515015. N Engl J Med. 2016. PMID: 26863367 No abstract available.
Similar articles
-
A Leadless Intracardiac Transcatheter Pacing System.N Engl J Med. 2016 Feb 11;374(6):533-41. doi: 10.1056/NEJMoa1511643. Epub 2015 Nov 9. N Engl J Med. 2016. PMID: 26551877 Clinical Trial.
-
A Dual-Chamber Leadless Pacemaker.N Engl J Med. 2023 Jun 22;388(25):2360-2370. doi: 10.1056/NEJMoa2300080. Epub 2023 May 20. N Engl J Med. 2023. PMID: 37212442
-
Permanent leadless cardiac pacing: results of the LEADLESS trial.Circulation. 2014 Apr 8;129(14):1466-71. doi: 10.1161/CIRCULATIONAHA.113.006987. Epub 2014 Mar 24. Circulation. 2014. PMID: 24664277 Clinical Trial.
-
Permanent Leadless Cardiac Pacemaker Therapy: A Comprehensive Review.Circulation. 2017 Apr 11;135(15):1458-1470. doi: 10.1161/CIRCULATIONAHA.116.025037. Circulation. 2017. PMID: 28396380 Review.
-
Leadless Cardiac Pacemakers: Pacing Paradigm Change.Curr Cardiol Rep. 2015 Aug;17(8):68. doi: 10.1007/s11886-015-0619-3. Curr Cardiol Rep. 2015. PMID: 26233700 Review.
Cited by
-
Double Perclose increases the efficiency of leadless pacemaker implantation: A propensity score-matched analysis.Heart Rhythm O2. 2024 Aug 9;5(10):750-753. doi: 10.1016/j.hroo.2024.07.018. eCollection 2024 Oct. Heart Rhythm O2. 2024. PMID: 39524051 Free PMC article. No abstract available.
-
A rate-responsive duty-cycling protocol for leadless pacemaker synchronization.Biomed Eng Lett. 2024 Aug 19;14(6):1397-1407. doi: 10.1007/s13534-024-00413-z. eCollection 2024 Nov. Biomed Eng Lett. 2024. PMID: 39465113 Free PMC article.
-
Development of Pericarditis Following Implantation of Micra Leadless Pacemaker.Cureus. 2024 Sep 23;16(9):e70027. doi: 10.7759/cureus.70027. eCollection 2024 Sep. Cureus. 2024. PMID: 39449952 Free PMC article.
-
Transcatheter non-acute retrieval of the tine-based leadless ventricular pacemaker.Europace. 2024 Oct 3;26(10):euae256. doi: 10.1093/europace/euae256. Europace. 2024. PMID: 39374848 Free PMC article.
-
Percutaneous retrieval of an atrial leadless pacemaker from the left ventricular outflow tract.HeartRhythm Case Rep. 2024 Jun 21;10(9):651-655. doi: 10.1016/j.hrcr.2024.06.012. eCollection 2024 Sep. HeartRhythm Case Rep. 2024. PMID: 39355820 Free PMC article. No abstract available.
Publication types
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous