Validation of the Oncentra Brachy Advanced Collapsed cone Engine for a commercial (192)Ir source using heterogeneous geometries

Brachytherapy. 2015 Nov-Dec;14(6):939-52. doi: 10.1016/j.brachy.2015.08.003. Epub 2015 Sep 26.

Abstract

Purpose: To validate the Advanced Collapsed cone Engine (ACE) dose calculation engine of Oncentra Brachy (OcB) treatment planning system using an (192)Ir source.

Methods and materials: Two levels of validation were performed, conformant to the model-based dose calculation algorithm commissioning guidelines of American Association of Physicists in Medicine TG-186 report. Level 1 uses all-water phantoms, and the validation is against TG-43 methodology. Level 2 uses real-patient cases, and the validation is against Monte Carlo (MC) simulations. For each case, the ACE and TG-43 calculations were performed in the OcB treatment planning system. ALGEBRA MC system was used to perform MC simulations.

Results: In Level 1, the ray effect depends on both accuracy mode and the number of dwell positions. The volume fraction with dose error ≥2% quickly reduces from 23% (13%) for a single dwell to 3% (2%) for eight dwell positions in the standard (high) accuracy mode. In Level 2, the 10% and higher isodose lines were observed overlapping between ACE (both standard and high-resolution modes) and MC. Major clinical indices (V100, V150, V200, D90, D50, and D2cc) were investigated and validated by MC. For example, among the Level 2 cases, the maximum deviation in V100 of ACE from MC is 2.75% but up to ~10% for TG-43. Similarly, the maximum deviation in D90 is 0.14 Gy between ACE and MC but up to 0.24 Gy for TG-43.

Conclusion: ACE demonstrated good agreement with MC in most clinically relevant regions in the cases tested. Departure from MC is significant for specific situations but limited to low-dose (<10% isodose) regions.

Keywords: ACE; Advanced Collapsed cone Engine; Brachytherapy; Collapsed-cone convolution; Dose calculation.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Algorithms
  • Brachytherapy*
  • Humans
  • Iridium Radioisotopes / therapeutic use*
  • Monte Carlo Method
  • Phantoms, Imaging
  • Radiotherapy Dosage*
  • Radiotherapy Planning, Computer-Assisted / methods*

Substances

  • Iridium Radioisotopes