Correlation between Serum Levels of 3,3',5'-Triiodothyronine and Thyroid Hormones Measured by Liquid Chromatography-Tandem Mass Spectrometry and Immunoassay

PLoS One. 2015 Oct 1;10(10):e0138864. doi: 10.1371/journal.pone.0138864. eCollection 2015.

Abstract

Objective: For measuring serum 3,3',5'-triiodothyronine (rT3) levels, radioimmunoassay (RIA) has traditionally been used owing to the lack of other reliable methods; however, it has recently become difficult to perform. Meanwhile, liquid chromatography-tandem mass spectrometry (LC-MS/MS) has recently been attracting attention as a novel alternative method in clinical chemistry. To the best of our knowledge, there are no studies to date comparing results of the quantification of human serum rT3 between LC-MS/MS and RIA. We therefore examined the feasibility of LC-MS/MS as a novel alternative method for measuring serum rT3, thyroxine (T4), and 3,5,3'-triiodothyronine (T3) levels.

Methods: Assay validation was performed by LC-MS/MS using quality control samples of rT3, T4, and T3 at 4 various concentrations which were prepared from reference compounds. Serum samples of 50 outpatients in our department were quantified both by LC-MS/MS and conventional immunoassay for rT3, T4, and T3. Correlation coefficients between the 2 measurement methods were statistically analyzed respectively.

Results: Matrix effects were not observed with our method. Intra-day and inter-day precisions were less than 10.8% and 9.6% for each analyte at each quality control level, respectively. Intra-day and inter-day accuracies were between 96.2% and 110%, and between 98.3% and 108.6%, respectively. The lower limit of quantification was 0.05 ng/mL. Strong correlations were observed between the 2 measurement methods (correlation coefficient, T4: 0.976, p < 0.001; T3: 0.912, p < 0.001; rT3: 0.928, p < 0.001).

Conclusions: Our LC-MS/MS system requires no manual cleanup operation, and the process after application of a sample is fully automated; furthermore, it was found to be highly sensitive, and superior in both precision and accuracy. The correlation between the 2 methods over a wide range of concentrations was strong. LC-MS/MS is therefore expected to become a useful tool for clinical diagnosis and research.

MeSH terms

  • Blood Chemical Analysis / methods*
  • Chromatography, Liquid / methods*
  • Feasibility Studies
  • Female
  • Humans
  • Linear Models
  • Male
  • Middle Aged
  • Radioimmunoassay / methods*
  • Reproducibility of Results
  • Tandem Mass Spectrometry / methods*
  • Thyroxine / blood
  • Triiodothyronine / blood
  • Triiodothyronine, Reverse / blood*

Substances

  • Triiodothyronine
  • Triiodothyronine, Reverse
  • Thyroxine

Grants and funding

The authors declare that ASKA Pharmaceutical Co., Ltd. provided support in the form of salaries for authors HN and TI, but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the "author contributions" section.