Improving Optoacoustic Image Quality via Geometric Pixel Super-Resolution Approach

IEEE Trans Med Imaging. 2016 Mar;35(3):812-8. doi: 10.1109/TMI.2015.2497159. Epub 2015 Nov 3.

Abstract

High fidelity optoacoustic (photoacoustic) tomography requires dense spatial sampling of optoacoustic signals using point acoustic detectors. However, in practice, spatial resolution of the images is often limited by limited sampling either due to coarse multi-element arrays or time in raster scan measurements. Herein, we investigate a method that integrates information from multiple optoacoustic images acquired at sub-diffraction steps into one high resolution image by means of an iterative registration algorithm. Experimental validations performed in target phantoms and ex vivo tissue samples confirm that the suggested approach renders significant improvements in terms of optoacoustic image resolution and quality without introducing significant alterations into the signal acquisition hardware or inversion algorithms.