Accurate Quantification Methods for Aortic Insufficiency Severity in Patients With LVAD: Role of Diastolic Flow Acceleration and Systolic-to-Diastolic Peak Velocity Ratio of Outflow Cannula
- PMID: 26684975
- DOI: 10.1016/j.jcmg.2015.06.020
Accurate Quantification Methods for Aortic Insufficiency Severity in Patients With LVAD: Role of Diastolic Flow Acceleration and Systolic-to-Diastolic Peak Velocity Ratio of Outflow Cannula
Abstract
Objectives: The aim of this study was to develop a technique to measure regurgitant flow throughout the entire cardiac cycle that would more accurately measure aortic insufficiency (AI) severity in patients with continuous-flow left ventricular assist devices (CF-LVADs).
Background: AI is a common problem after CF-LVAD implantation. Current echocardiographic evaluation of AI does not take into account the unique flow properties present in patients with CF-LVADs.
Methods: In this prospective study, patients with LVADs who had varying degrees of AI (N = 20) underwent simultaneous right-sided heart catheterization (RHC) and transthoracic echocardiography (TTE). Regurgitant fraction (RF) across the aortic valve was calculated by subtracting the cardiac output obtained using the Fick method from the total systemic flow measured using the sum of the product of the velocity time integral and the cross-sectional area of the LVAD outflow cannula and aortic valve, respectively. The RFs were then compared with the following: 1) traditional TTE grading parameters; and 2) new TTE parameters unique to LVAD physiology, namely the diastolic flow acceleration and the systolic-to-diastolic peak velocity (S/D) ratio of the LVAD outflow cannula.
Results: Patients without evidence of AI had an RF approaching zero (2.4 ± 4.6%). Patients with trace and mild AI had an RF of 31.0 ± 5.4%, whereas patients with moderate or severe AI had an RF of 45.8 ± 3.6%. RF correlated better with pulmonary capillary wedge pressure (PCWP) than with vena contracta (correlation coefficient [R] = 0.73 vs. 0.56). The new TTE parameters (S/D ratio and diastolic acceleration) highly correlated with RF (R = 0.91 and 0.94, respectively) and more strongly correlated with PCWP than did vena contracta (R = 0.82 and 0.65 vs. 0.56).
Conclusions: RF measured by simultaneous RHC and TTE better correlates with clinical filling pressures than do traditional TTE parameters and may identify significant AI that could be underestimated using conventional measures. Novel TTE parameters, unique to CF-LVAD physiology, better correlate with RF and filling pressures than do our current TTE measurements.
Keywords: LVAD; aortic insufficiency; heart failure.
Copyright © 2016 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Comment in
-
Dealing With Unintended Consequences: Continuous-Flow LVADs and Aortic Insufficiency.JACC Cardiovasc Imaging. 2016 Jun;9(6):652-4. doi: 10.1016/j.jcmg.2015.07.016. Epub 2015 Dec 9. JACC Cardiovasc Imaging. 2016. PMID: 26684965 No abstract available.
Similar articles
-
Novel echocardiographic parameters of aortic insufficiency in continuous-flow left ventricular assist devices and clinical outcome.J Heart Lung Transplant. 2016 Aug;35(8):976-85. doi: 10.1016/j.healun.2016.05.009. Epub 2016 May 20. J Heart Lung Transplant. 2016. PMID: 27373822 Free PMC article.
-
Percutaneous transcatheter aortic valve closure successfully treats left ventricular assist device-associated aortic insufficiency and improves cardiac hemodynamics.JACC Cardiovasc Interv. 2013 Jan;6(1):84-9. doi: 10.1016/j.jcin.2012.08.021. JACC Cardiovasc Interv. 2013. PMID: 23347865
-
Efficacy and durability of central oversewing for treatment of aortic insufficiency in patients with continuous-flow left ventricular assist devices.J Heart Lung Transplant. 2014 Sep;33(9):937-42. doi: 10.1016/j.healun.2014.04.017. Epub 2014 May 9. J Heart Lung Transplant. 2014. PMID: 24997496
-
Management of aortic insufficiency in the continuous flow left ventricular assist device population.Curr Heart Fail Rep. 2014 Mar;11(1):103-10. doi: 10.1007/s11897-013-0172-6. Curr Heart Fail Rep. 2014. PMID: 24193452 Review.
-
Comprehensive assessment of the aortic valve in critically ill patients for the non-cardiologist. Part II: Chronic aortic regurgitation of the native valve.Anaesthesiol Intensive Ther. 2021;53(1):55-68. doi: 10.5114/ait.2021.104892. Anaesthesiol Intensive Ther. 2021. PMID: 33788504 Free PMC article. Review.
Cited by
-
Aortic valve disorders and left ventricular assist devices.Front Cardiovasc Med. 2023 Feb 23;10:1098348. doi: 10.3389/fcvm.2023.1098348. eCollection 2023. Front Cardiovasc Med. 2023. PMID: 36910539 Free PMC article. Review.
-
Patient-specific, echocardiography compatible flow loop model of aortic valve regurgitation in the setting of a mechanical assist device.Front Cardiovasc Med. 2023 Feb 8;10:994431. doi: 10.3389/fcvm.2023.994431. eCollection 2023. Front Cardiovasc Med. 2023. PMID: 36844719 Free PMC article.
-
Key questions about aortic insufficiency in patients with durable left ventricular assist devices.Front Cardiovasc Med. 2022 Nov 24;9:1068707. doi: 10.3389/fcvm.2022.1068707. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 36505355 Free PMC article. Review.
-
Prediction, prevention, and management of right ventricular failure after left ventricular assist device implantation: A comprehensive review.Front Cardiovasc Med. 2022 Nov 3;9:1040251. doi: 10.3389/fcvm.2022.1040251. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 36407460 Free PMC article. Review.
-
A computational study of aortic insufficiency in patients supported with continuous flow left ventricular assist devices: Is it time for a paradigm shift in management?Front Cardiovasc Med. 2022 Oct 20;9:933321. doi: 10.3389/fcvm.2022.933321. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 36337891 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources