Mechanical deformation of thin film platinum under electrical stimulation

Annu Int Conf IEEE Eng Med Biol Soc. 2015 Aug:2015:1045-8. doi: 10.1109/EMBC.2015.7318544.

Abstract

Thin-film-based electrodes used to interact with nervous tissue often fail quickly if used for electrical stimulation, impairing their translation into long-term clinical applications. We initiated investigations about the mechanical load on thin-film electrodes caused by the fact of electrical stimulation. Platinum electrodes of Ø 300μm on a polyimide carrier were subjected to approximately 50 000 asymmetrical, biphasic stimulation pulses in vitro. The electrode's surface was investigated optically by means of white-light interferometry. The structural expansion for the metallic surface subjected to stimulation was measured to reach roughly 30%. The study points towards a failure mechanism of thin-films being of mechanical nature, inherent to the unavoidable electrochemical processes involved (change in lattice constants) during electrical stimulation at the electrode's surface. Based on further scientific facts, we set 3 hypotheses for the exact mechanisms involved in the failure of thin-films used for electrical stimulation, opening a new door for research and improvement of novel neuroprosthetic devices.

MeSH terms

  • Electric Stimulation Therapy
  • Electric Stimulation*
  • Electrodes
  • Platinum

Substances

  • Platinum