The effect of jet shape on jet injection

Annu Int Conf IEEE Eng Med Biol Soc. 2015:2015:7350-3. doi: 10.1109/EMBC.2015.7320089.

Abstract

The effects of the dispersion pattern of a needle-free jet injector are explored. The shape of the jets were compared using a high-speed video camera and jet injections of collimated and dispersed fluid jets with a Lorentz-force actuated jet injector were made into acrylamide gel and post-mortem porcine tissue. A custom-built high-speed X-ray imaging system was used in order to observe the dynamics of the dispersion mechanism for each injection in real time. We show that a collimated jet stream results in greater tissue penetration than a dispersed jet stream.