First clinical application of a prompt gamma based in vivo proton range verification system

Radiother Oncol. 2016 Feb;118(2):232-7. doi: 10.1016/j.radonc.2016.01.004. Epub 2016 Jan 13.

Abstract

Background and purpose: To improve precision of particle therapy, in vivo range verification is highly desirable. Methods based on prompt gamma rays emitted during treatment seem promising but have not yet been applied clinically. Here we report on the worldwide first clinical application of prompt gamma imaging (PGI) based range verification.

Material and methods: A prototype of a knife-edge shaped slit camera was used to measure the prompt gamma ray depth distribution during a proton treatment of a head and neck tumor for seven consecutive fractions. Inter-fractional variations of the prompt gamma profile were evaluated. For three fractions, in-room control CTs were acquired and evaluated for dose relevant changes.

Results: The measurement of PGI profiles during proton treatment was successful. Based on the PGI information, inter-fractional global range variations were in the range of ±2 mm for all evaluated fractions. This is in agreement with the control CT evaluation showing negligible range variations of about 1.5mm.

Conclusions: For the first time, range verification based on prompt gamma imaging was applied for a clinical proton treatment. With the translation from basic physics experiments into clinical operation, the potential to improve the precision of particle therapy with this technique has increased considerably.

Keywords: First in man; Prompt gamma imaging; Proton therapy; Range verification; Slit camera.

Publication types

  • Case Reports
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Carcinoma, Adenoid Cystic / diagnostic imaging
  • Carcinoma, Adenoid Cystic / radiotherapy*
  • Gamma Rays / therapeutic use*
  • Head and Neck Neoplasms / diagnostic imaging
  • Head and Neck Neoplasms / radiotherapy*
  • Humans
  • Male
  • Middle Aged
  • Proton Therapy / methods*
  • Tomography, X-Ray Computed