Evaluation of Liquid-Based Swab Transport Systems against the New Approved CLSI M40-A2 Standard

J Clin Microbiol. 2016 Apr;54(4):1152-6. doi: 10.1128/JCM.03337-15. Epub 2016 Feb 3.

Abstract

Following revised information pertaining to newer swab types and testing protocols in the new CLSI M40-A2 standard, we evaluated three liquid swab transport systems for the recovery of aerobic, anaerobic, and fastidious organisms at room temperature and at 4°C. All tested liquid swab transport systems were fully compliant with the M40-A2 standard, with acceptable performance at both temperatures after the full specified holding period, using both qualitative (roll-plate) and quantitative (swab elution) methods.

MeSH terms

  • Bacteria, Aerobic / isolation & purification*
  • Bacteria, Anaerobic / isolation & purification*
  • Bacteriological Techniques / methods*
  • Culture Media*
  • Humans
  • Sensitivity and Specificity
  • Specimen Handling / methods*

Substances

  • Culture Media