Quantification of dental prostheses on cone-beam CT images by the Taguchi method

J Appl Clin Med Phys. 2016 Jan 8;17(1):207-220. doi: 10.1120/jacmp.v17i1.5826.

Abstract

The gray values accuracy of dental cone-beam computed tomography (CBCT) is affected by dental metal prostheses. The distortion of dental CBCT gray values could lead to inaccuracies of orthodontic and implant treatment. The aim of this study was to quantify the effect of scanning parameters and dental metal prostheses on the accuracy of dental cone-beam computed tomography (CBCT) gray values using the Taguchi method. Eight dental model casts of an upper jaw including prostheses, and a ninth prosthesis-free dental model cast, were scanned by two dental CBCT devices. The mean gray value of the selected circular regions of interest (ROIs) were measured using dental CBCT images of eight dental model casts and were compared with those measured from CBCT images of the prosthesis-free dental model cast. For each image set, four consecutive slices of gingiva were selected. The seven factors (CBCTs, occlusal plane canting, implant connection, prosthesis position, coping material, coping thickness, and types of dental restoration) were used to evaluate scanning parameter and dental prostheses effects. Statistical methods of signal to noise ratio (S/N) and analysis of variance (ANOVA) with 95% confidence were applied to quantify the effects of scanning parameters and dental prostheses on dental CBCT gray values accuracy. For ROIs surrounding dental prostheses, the accuracy of CBCT gray values were affected primarily by implant connection (42%), followed by type of restoration (29%), prostheses position (19%), coping material (4%), and coping thickness (4%). For a single crown prosthesis (without support of implants) placed in dental model casts, gray value differences for ROIs 1-9 were below 12% and gray value differences for ROIs 13-18 away from pros-theses were below 10%. We found the gray value differences set to be between 7% and 8% for regions next to a single implant-supported titanium prosthesis, and between 46% and 59% for regions between double implant-supported, nickel-chromium alloys (Ni-Cr) prostheses. Quantification of the effect of prostheses and scanning parameters on dental CBCT gray values was assessed.

MeSH terms

  • Cone-Beam Computed Tomography / methods*
  • Dental Implants
  • Dental Prosthesis*
  • Gingiva / diagnostic imaging*
  • Gingiva / pathology
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Models, Anatomic*

Substances

  • Dental Implants