Verification of I-125 brachytherapy source strength for use in radioactive seed localization procedures

Appl Radiat Isot. 2016 Jun:112:62-8. doi: 10.1016/j.apradiso.2016.03.006. Epub 2016 Mar 15.

Abstract

A general-purpose nuclear medicine dose calibrator was assessed as a potential replacement for a dedicated air-communicating well-type ionization chamber (brachytherapy source strength verification instrument) for (125)I seed source strength verification for radioactive seed localization, where less stringent accuracy tolerances may be acceptable. The accuracy, precision and reproducibility of the dose calibrator were measured and compared to regulatory requirements. The results of this work indicate that a dose calibrator can be used for (125)I seed source strength verification for radioactive seed localization.

Keywords: Dose calibrator; Radioactive seed localization; Source strength verification.