Development and validation of a correction equation for Corvis tonometry
- PMID: 27049961
- DOI: 10.1080/10255842.2015.1077515
Development and validation of a correction equation for Corvis tonometry
Abstract
Primary objective: This study uses numerical analysis and validation against clinical data to develop a method to correct intraocular pressure (IOP) measurements obtained using the Corvis Tonometer for the effects of central corneal thickness (CCT), and age.
Materials and methods: Finite element analysis was conducted to simulate the effect of tonometric air pressure on the intact eye globe. The analyses considered eyes with wide variations in IOP (10-30 mm Hg), CCT (445-645 microns), R (7.2-8.4 mm), shape factor, P (0.6-1) and age (30-90 years). In each case, corneal deformation was predicted and used to estimate the IOP measurement by Corvis (CVS-IOP). Analysis of the results led to an algorithm relating estimates of true IOP as a function of CVS-IOP, CCT and age. All other parameters had negligible effect on CVS-IOP and have therefore been omitted from the algorithm. Predictions of corrected CVS-IOP, as obtained by applying the algorithm to a clinical data-set involving 634 eyes, were assessed for their association with the cornea stiffness parameters; CCT and age.
Results: Analysis of CVS-IOP measurements within the 634-large clinical data-set showed strong correlation with CCT (3.06 mm Hg/100 microns, r(2) = 0.204) and weaker correlation with age (0.24 mm Hg/decade, r(2) = 0.009). Applying the algorithm to IOP measurements resulted in IOP estimations that became less correlated with both CCT (0.04 mm Hg/100 microns, r(2) = 0.005) and age (0.09 mm Hg/decade, r(2) = 0.002).
Conclusions: The IOP correction process developed in this study was successful in reducing reliance of IOP measurements on both corneal thickness and age in a healthy European population.
Keywords: Corvis ST; Tonometry; cornea; intraocular pressure; ocular biomechanics.
Similar articles
-
Clinical Evaluation of Methods to Correct Intraocular Pressure Measurements by the Goldmann Applanation Tonometer, Ocular Response Analyzer, and Corvis ST Tonometer for the Effects of Corneal Stiffness Parameters.J Glaucoma. 2016 Jun;25(6):510-9. doi: 10.1097/IJG.0000000000000359. J Glaucoma. 2016. PMID: 26709500
-
Ex-vivo experimental validation of biomechanically-corrected intraocular pressure measurements on human eyes using the CorVis ST.Exp Eye Res. 2018 Oct;175:98-102. doi: 10.1016/j.exer.2018.06.013. Epub 2018 Jun 15. Exp Eye Res. 2018. PMID: 29908883
-
Evaluation of a novel Scheimpflug-based non-contact tonometer in healthy subjects and patients with ocular hypertension and glaucoma.Br J Ophthalmol. 2013 Nov;97(11):1410-4. doi: 10.1136/bjophthalmol-2013-303400. Epub 2013 Aug 22. Br J Ophthalmol. 2013. PMID: 23969314
-
Corneal Deformation Parameters Provided by the Corvis-ST Pachy-Tonometer in Healthy Subjects and Glaucoma Patients.J Glaucoma. 2015 Oct-Nov;24(8):568-74. doi: 10.1097/IJG.0000000000000133. J Glaucoma. 2015. PMID: 25318572 Clinical Trial.
-
Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.Surv Ophthalmol. 2000 Mar-Apr;44(5):367-408. doi: 10.1016/s0039-6257(00)00110-7. Surv Ophthalmol. 2000. PMID: 10734239 Review.
Cited by
-
Scheimpflug-Based Corneal Biomechanical Analysis As A Predictor of Glaucoma in Eyes With High Myopia.Clin Ophthalmol. 2024 Feb 22;18:545-563. doi: 10.2147/OPTH.S426635. eCollection 2024. Clin Ophthalmol. 2024. PMID: 38410632 Free PMC article.
-
Biomechanical properties measured with dynamic Scheimpflug analyzer in central serous chorioretinopathy.Graefes Arch Clin Exp Ophthalmol. 2024 Jan 29. doi: 10.1007/s00417-024-06378-0. Online ahead of print. Graefes Arch Clin Exp Ophthalmol. 2024. PMID: 38285248
-
Comparing corneal biomechanic changes among solo cataract surgery, microhook ab interno trabeculotomy and iStent implantation.Sci Rep. 2023 Nov 6;13(1):19148. doi: 10.1038/s41598-023-46709-5. Sci Rep. 2023. PMID: 37932377 Free PMC article.
-
The influence of topical anesthetic and fluorescein on non-contact tonometry measurements using ultra-high-speed dynamic Scheimpflug.Sci Rep. 2023 Oct 19;13(1):17864. doi: 10.1038/s41598-023-45165-5. Sci Rep. 2023. PMID: 37857798 Free PMC article.
-
Assessment of age-related change of the ocular support system.Front Bioeng Biotechnol. 2023 Jul 10;11:1146828. doi: 10.3389/fbioe.2023.1146828. eCollection 2023. Front Bioeng Biotechnol. 2023. PMID: 37492801 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources