Low-dose abdominal computed tomography for detection of urinary stone disease - Impact of additional spectral shaping of the X-ray beam on image quality and dose parameters

Eur J Radiol. 2016 Jun;85(6):1058-62. doi: 10.1016/j.ejrad.2016.03.016. Epub 2016 Mar 22.

Abstract

Objectives: To evaluate a novel tin filter-based abdominal CT protocol for urolithiasis in terms of image quality and CT dose parameters.

Methods: 130 consecutive patients with suspected urolithiasis underwent non-enhanced CT with three different protocols: 48 patients (group 1) were examined at tin-filtered 150kV (150kV Sn) on a third-generation dual-source-CT, 33 patients were examined with automated kV-selection (110-140kV) based on the scout view on the same CT-device (group 2), and 49 patients were examined on a second-generation dual-source-CT (group 3) with automated kV-selection (100-140kV). Automated exposure control was active in all groups. Image quality was subjectively evaluated on a 5-point-likert-scale by two radiologists and interobserver agreement as well as signal-to-noise-ratio (SNR) was calculated. Dose-length-product (DLP) and volume CT dose index (CTDIvol) were compared.

Results: Image quality was rated in favour for the tin filter protocol with excellent interobserver agreement (ICC=0.86-0.91) and the difference reached statistical significance (p<0.001). SNR was significantly higher in group 1 and 2 compared to second-generation DSCT (p<0.001). On third-generation dual-source CT, there was no significant difference in SNR between the 150kV Sn and the automated kV selection protocol (p=0.5). The DLP of group 1 was 23% and 21% (p<0.002) lower in comparison to group 2 and 3, respectively. So was the CTDIvol of group 1 compared to group 2 (-36%) and 3 (-32%) (p<0.001).

Conclusion: Additional shaping of a 150kV source spectrum by a tin filter substantially lowers patient exposure while improving image quality on un-enhanced abdominal computed tomography for urinary stone disease.

Keywords: CT; Low-dose abdominal CT; Spectral shaping; Tin filter; Urinary stone disease.

Publication types

  • Comparative Study

MeSH terms

  • Adipose Tissue / diagnostic imaging
  • Adult
  • Artifacts
  • Female
  • Filtration / instrumentation
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Male
  • Middle Aged
  • Observer Variation
  • Psoas Muscles / diagnostic imaging
  • Radiation Dosage*
  • Radiographic Image Enhancement / instrumentation
  • Radiographic Image Enhancement / methods*
  • Radiography, Abdominal / instrumentation
  • Radiography, Abdominal / methods*
  • Signal-To-Noise Ratio
  • Tin
  • Tomography, X-Ray Computed / instrumentation
  • Tomography, X-Ray Computed / methods*
  • Urinary Calculi / diagnostic imaging*

Substances

  • Tin