Utility of Decision Rules for Transcutaneous Bilirubin Measurements
- PMID: 27244792
- PMCID: PMC4845868
- DOI: 10.1542/peds.2015-3032
Utility of Decision Rules for Transcutaneous Bilirubin Measurements
Abstract
Background: Transcutaneous bilirubin (TcB) meters are widely used for screening newborns for jaundice, with a total serum bilirubin (TSB) measurement indicated when the TcB value is classified as "positive" by using a decision rule. The goal of our study was to assess the clinical utility of 3 recommended TcB screening decision rules.
Methods: Paired TcB/TSB measurements were collected at 34 newborn nursery sites. At 27 sites (sample 1), newborns were routinely screened with a TcB measurement. For sample 2, sites that typically screen with TSB levels also obtained a TcB measurement for the study. Three decision rules to define a positive TcB measurement were evaluated: ≥75th percentile on the Bhutani nomogram, 70% of the phototherapy level, and within 3 mg/dL of the phototherapy threshold. The primary outcome was a TSB level at/above the phototherapy threshold. The rate of false-negative TcB screens and percentage of blood draws avoided were calculated for each decision rule.
Results: For sample 1, data were analyzed on 911 paired TcB-TSB measurements from a total of 8316 TcB measurements. False-negative rates were <10% with all decision rules; none identified all 31 newborns with a TSB level at/above the phototherapy threshold. The percentage of blood draws avoided ranged from 79.4% to 90.7%. In sample 2, each rule correctly identified all 8 newborns with TSB levels at/above the phototherapy threshold.
Conclusions: Although all of the decision rules can be used effectively to screen newborns for jaundice, each will "miss" some infants with a TSB level at/above the phototherapy threshold.
Copyright © 2016 by the American Academy of Pediatrics.
Conflict of interest statement
Figures
Similar articles
-
Discrepancies between transcutaneous and serum bilirubin measurements.Pediatrics. 2015 Feb;135(2):224-31. doi: 10.1542/peds.2014-1919. Epub 2015 Jan 19. Pediatrics. 2015. PMID: 25601981 Free PMC article.
-
Transcutaneous bilirubin levels in newborns <35 weeks' gestation.J Perinatol. 2015 Sep;35(9):739-44. doi: 10.1038/jp.2015.34. Epub 2015 Jun 25. J Perinatol. 2015. PMID: 26110497
-
Validation of transcutaneous bilirubinometry during phototherapy for detection and monitoring of neonatal jaundice in a low-income setting.Paediatr Int Child Health. 2020 Feb;40(1):25-29. doi: 10.1080/20469047.2019.1598126. Epub 2019 Apr 11. Paediatr Int Child Health. 2020. PMID: 30973082
-
Effect of Phototherapy on the Reliability of Transcutaneous Bilirubin Devices in Term and Near-Term Infants: A Systematic Review and Meta-Analysis.Neonatology. 2016;109(3):203-12. doi: 10.1159/000442195. Epub 2016 Jan 21. Neonatology. 2016. PMID: 26789390 Review.
-
Transcutaneous bilirubinometry versus total serum bilirubin measurement for newborns.Cochrane Database Syst Rev. 2023 May 9;5(5):CD012660. doi: 10.1002/14651858.CD012660.pub2. Cochrane Database Syst Rev. 2023. PMID: 37158489 Free PMC article. Review.
Cited by
-
Early-Term Neonates Demonstrate a Higher Likelihood of Requiring Phototherapy Compared to Those Born Full-Term.Children (Basel). 2023 Nov 16;10(11):1819. doi: 10.3390/children10111819. Children (Basel). 2023. PMID: 38002910 Free PMC article.
-
Self-Powered Wearable Biosensor in a Baby Diaper for Monitoring Neonatal Jaundice through a Hydrovoltaic-Biosensing Coupling Effect of ZnO Nanoarray.Biosensors (Basel). 2022 Mar 6;12(3):164. doi: 10.3390/bios12030164. Biosensors (Basel). 2022. PMID: 35323434 Free PMC article.
-
Simplifying Hyperbilirubinemia Risk Estimation.Pediatrics. 2021 May;147(5):e2020046284. doi: 10.1542/peds.2020-046284. Pediatrics. 2021. PMID: 33903165 Free PMC article. No abstract available.
-
Use of a Smartphone App to Assess Neonatal Jaundice.Pediatrics. 2017 Sep;140(3):e20170312. doi: 10.1542/peds.2017-0312. Pediatrics. 2017. PMID: 28842403 Free PMC article.
References
-
- Maisels MJ, Bhutani VK, Bogen D, Newman TB, Stark AR, Watchko JF. Hyperbilirubinemia in the newborn infant > or =35 weeks’ gestation: an update with clarifications. Pediatrics. 2009;124(4):1193–1198 - PubMed
-
- Engle WD, Jackson GL, Sendelbach D, Manning D, Frawley WH. Assessment of a transcutaneous device in the evaluation of neonatal hyperbilirubinemia in a primarily Hispanic population. Pediatrics. 2002;110(1 pt 1):61–67 - PubMed
-
- Engle WD, Jackson GL, Stehel EK, Sendelbach DM, Manning MD. Evaluation of a transcutaneous jaundice meter following hospital discharge in term and near-term neonates. J Perinatol. 2005;25(7):486–490 - PubMed
-
- Bhutani VK, Gourley GR, Adler S, Kreamer B, Dalin C, Johnson LH. Noninvasive measurement of total serum bilirubin in a multiracial predischarge newborn population to assess the risk of severe hyperbilirubinemia. Pediatrics. 2000;106(2). Available at: www.pediatrics.org/cgi/content/full/106/2/E17 - PubMed