Observational Study
. 2016 Sep;50:1-4.
doi: 10.1016/j.cct.2016.07.007. Epub 2016 Jul 6.

Results of a Near Continuous Glucose Monitoring Technology in Surgical Intensive Care and Trauma

Affiliations
Observational Study

Results of a Near Continuous Glucose Monitoring Technology in Surgical Intensive Care and Trauma

Eden Nohra et al. Contemp Clin Trials. .

Abstract

Introduction: Near-continuous glucose monitoring is expected to increase time in range (TIR) of 80-120mg/dL and to avoid hypoglycemia without increasing workload. We investigated a near-continuous glucose monitor in surgical critically ill and trauma patients.

Methods: Patients were enrolled at a surgical intensive care unit associated with a level 1 trauma center. Glucose measurements were compared to the gold standard Yellow Springs Instrument (YSI). The technology withdraws 0.13mL of blood every 15min from a central venous line, centrifuges the sample, and uses mid-infrared spectroscopy to measure glucose. We plotted a Clarke Error Grid, calculated Mean Absolute Relative Deviation (MARD) to analyze trend accuracy, and we present a Bland Altman plot of device versus standard glucose measurements.

Results: 24 patients were enrolled. One patient was withdrawn due to poor blood return from central venous line. A total of 347 glucose measurements from 23 patients were compared to the gold standard. 94.8% of the data points were in zone A of the Clarke Error Grid and 5.2% in zone B. The MARD was 8.02%. The majority of data points achieved the benchmark for accuracy. The remaining 5.2% are clinically benign. The MARD was below 10%. The Bland Altman plot shows good agreement between the device and reference glucose measurements. There were no device related adverse events.

Conclusion: Our data suggests that near continuous monitoring via infrared spectroscopy is safe and accurate for use in critically ill surgical and trauma patients. A large scale multi-center study is underway to confirm these findings.

Trial registration: ClinicalTrials.gov NCT02211300.

Keywords: Accuracy; Blood glucose meter; Continuous glucose monitoring; Glucose control; Near-infrared spectroscopy; Surgical intensive care.