Measurement of titanium in hip-replacement patients by inductively coupled plasma optical emission spectroscopy

Ann Clin Biochem. 2017 May;54(3):362-369. doi: 10.1177/0004563216662292. Epub 2016 Jul 15.

Abstract

Background Patients with metal-on-metal hip replacements require testing for cobalt and chromium. There may also be a need to test for titanium, which is used in the construction of the femoral stem in total hip replacements. It is not possible to use quadrupole inductively coupled plasma mass spectrometry due to interferences. Methods Titanium was measured using inductively coupled plasma optical emission spectroscopy using the emission line at 336.1 nm and Y (internal standard) at 371.0 nm. Internal quality control materials were prepared for blood and serum and concentrations assigned using a sector field-inductively coupled plasma mass spectrometer. A candidate whole blood certified reference material was also evaluated. Results The method had detection and quantitation limits of 0.6 and 1.9 µg/L, respectively. The respective bias (%) and measurement uncertainty ( U) (k = 2) were 3.3% and 2.0 µg/L (serum) and - 1.0% and 1.4 µg/L (whole blood). The respective repeatability and intermediate precision (%) were 5.1% and 10.9% (serum) and 2.4% and 8.6% (whole blood). The concentration of titanium was determined in patients' samples, serum (median = 2.4 µg/L, n = 897) and whole blood (median = 2.4 µg/L, n = 189). Serum is recommended for monitoring titanium in patients, since the concentration is higher than in whole blood and the matrix less problematic. In hip fluid samples, the concentrations were much higher (mean 58.5 µg/L, median 5.1 µg/L, n = 83). Conclusions A method based on inductively coupled plasma optical emission spectroscopy was developed and validated for measuring titanium in clinical samples.

Keywords: Total hip replacement; inductively coupled plasma optical emission spectroscopy; metal-on-metal; titanium.

Publication types

  • Validation Study

MeSH terms

  • Arthroplasty, Replacement, Hip*
  • Chromium / blood
  • Cobalt / blood
  • Female
  • Hip Prosthesis
  • Humans
  • Limit of Detection
  • Male
  • Quality Control
  • Reference Standards
  • Spectrophotometry, Atomic / standards*
  • Titanium / blood*

Substances

  • Chromium
  • Cobalt
  • Titanium