Impact of latest generation cardiac interventional X-ray equipment on patient image quality and radiation dose for trans-catheter aortic valve implantations

Br J Radiol. 2016 Nov;89(1067):20160269. doi: 10.1259/bjr.20160269. Epub 2016 Sep 29.

Abstract

Objectives: This study aimed to determine the impact on radiation dose and image quality of a new cardiac interventional X-ray system for trans-catheter aortic valve implantation (TAVI) patients compared to the previously-used cardiac X-ray system.

Methods: Patient dose and image data were retrospectively collected from a Philips AlluraClarity (new) and Siemens Axion Artis (reference) X-ray system. Patient dose area product (DAP) and fluoroscopy duration of 41 patient cases from each X-ray system were compared using a Wilcoxon test. Ten patient aortograms from each X-ray system were scored by 32 observers on a continuous scale to assess the clinical image quality at the given phase of the TAVI procedure. Scores were dichotomised by acceptability and analysed using a Chi-squared test.

Results: Significant reductions in patient dose (p << 0.001) were found for the new system with no significant change in fluoroscopy duration (p = 0.052); procedure DAP reduced by 55%, fluoroscopy DAP by 48% and "cine" acquisition DAP by 61%. There was no significant difference between image quality scores of the two X-ray systems (p = 0.06).

Conclusions: The new cardiac X-ray system demonstrated a very significant reduction in patient dose with no loss of clinical image quality. Advances in Knowledge: The huge growth of TAVI may impact on the radiation exposure of cardiac patients and particularly on operators including anaesthetists; cumulative exposure of interventional cardiologists performing high volume TAVI over 30-40 years may be harmful. The Phillips Clarity upgrade including improved image enhancement and optimised X-ray settings significantly reduced radiation without reducing clinically acceptable image quality.

Publication types

  • Comparative Study

MeSH terms

  • Fluoroscopy
  • Humans
  • Radiation Dosage
  • Radiography, Interventional / instrumentation*
  • Radiography, Thoracic / instrumentation*
  • Retrospective Studies
  • Risk Factors
  • Time Factors
  • Transcatheter Aortic Valve Replacement*