EUROSPINE 2016 FULL PAPER AWARD: Wire cerclage can restore the stability of the thoracic spine after median sternotomy: an in vitro study with entire rib cage specimens
- PMID: 27639711
- DOI: 10.1007/s00586-016-4768-x
EUROSPINE 2016 FULL PAPER AWARD: Wire cerclage can restore the stability of the thoracic spine after median sternotomy: an in vitro study with entire rib cage specimens
Abstract
Purpose: The influence of the anterior rib cage on the stability of the human thoracic spine is not completely known. One of the most common surgical interventions on the anterior rib cage is the longitudinal median sternotomy and its fixation by wire cerclage. Therefore, the purpose of this in vitro study was to examine, if wire cerclage can restore the stability of the human thoracic spine after longitudinal median sternotomy.
Methods: Six fresh frozen human thoracic spine specimens (C7-L1, 56 years in average, range 50-65), including the intact rib cage without intercostal muscles, were tested in a spinal loading simulator and monitored with an optical motion tracking system. While applying 2 Nm pure moment in flexion/extension (FE), lateral bending (LB), and axial rotation (AR), the range of motion (ROM) and neutral zone (NZ) of the functional spinal units of the thoracic spine (T1-T12) were studied (1) in intact condition, (2) after longitudinal median sternotomy, and (3) after sternal closure using wire cerclage.
Results: The longitudinal median sternotomy caused a significant increase of the thoracic spine ROM relative to the intact condition (FE: 12° ± 5°, LB: 18° ± 5°, AR: 25° ± 10°) in FE (+12 %) and AR (+22 %). As a result, the sagittal cut faces of the sternum slipped apart visibly. Wire cerclage fixation resulted in a significant decrease of the ROM in AR (-12 %) relative to condition after sternotomy. ROM increased relative to the intact condition, in AR even significantly (+8 %). The NZ showed a proportional behavior compared to the ROM in all loading planes, but it was distinctly higher in FE (72 %) and in LB (82 %) compared to the ROM than in AR (12 %).
Conclusions: In this in vitro study, the longitudinal median sternotomy resulted in a destabilization of the thoracic spine and relative motion of the sternal cut faces, which could be rectified by fixation with wire cerclage. However, the stability of the intact condition could not be reached. Nevertheless, a fixation of the sternum should be considered clinically to avoid instability of the spine and sternal pseudarthrosis.
Keywords: In vitro study; Longitudinal median sternotomy; Rib cage; Thoracic spine; Wire cerclage.
Similar articles
-
The rib cage stabilizes the human thoracic spine: An in vitro study using stepwise reduction of rib cage structures.PLoS One. 2017 Jun 1;12(6):e0178733. doi: 10.1371/journal.pone.0178733. eCollection 2017. PLoS One. 2017. PMID: 28570671 Free PMC article.
-
In vitro analysis of kinematics and elastostatics of the human rib cage during thoracic spinal movement for the validation of numerical models.J Biomech. 2019 Sep 20;94:147-157. doi: 10.1016/j.jbiomech.2019.07.041. Epub 2019 Aug 7. J Biomech. 2019. PMID: 31420155
-
The effect of follower load on the intersegmental coupled motion characteristics of the human thoracic spine: An in vitro study using entire rib cage specimens.J Biomech. 2018 Sep 10;78:36-44. doi: 10.1016/j.jbiomech.2018.06.025. Epub 2018 Jul 6. J Biomech. 2018. PMID: 30031651
-
Biomechanical analysis of screw constructs for atlantoaxial fixation in cadavers: a systematic review and meta-analysis.J Neurosurg Spine. 2015 Feb;22(2):151-61. doi: 10.3171/2014.10.SPINE13805. Epub 2014 Dec 5. J Neurosurg Spine. 2015. PMID: 25478824 Review.
-
Challenging the Conventional Standard for Thoracic Spine Range of Motion: A Systematic Review.JBJS Rev. 2016 Apr 5;4(4):e51-e511. doi: 10.2106/JBJS.RVW.O.00048. JBJS Rev. 2016. PMID: 27487429 Review.
Cited by
-
Postoperative C-reactive protein levels correlate with reduced spinal column mobility after median sternotomy: a prospective cohort study.J Thorac Dis. 2024 Jan 30;16(1):469-478. doi: 10.21037/jtd-23-1439. Epub 2024 Jan 12. J Thorac Dis. 2024. PMID: 38410601 Free PMC article.
-
[Biomechanics of thoracic wall instability].Unfallchirurgie (Heidelb). 2024 Mar;127(3):180-187. doi: 10.1007/s00113-023-01389-8. Epub 2023 Nov 14. Unfallchirurgie (Heidelb). 2024. PMID: 37964040 Review. German.
-
How Does the Rib Cage Affect the Biomechanical Properties of the Thoracic Spine? A Systematic Literature Review.Front Bioeng Biotechnol. 2022 Jun 15;10:904539. doi: 10.3389/fbioe.2022.904539. eCollection 2022. Front Bioeng Biotechnol. 2022. PMID: 35782518 Free PMC article.
-
Thoracic Spinal Stability and Motion Behavior Are Affected by the Length of Posterior Instrumentation After Vertebral Body Replacement, but Not by the Surgical Approach Type: An in vitro Study With Entire Rib Cage Specimens.Front Bioeng Biotechnol. 2020 Jun 9;8:572. doi: 10.3389/fbioe.2020.00572. eCollection 2020. Front Bioeng Biotechnol. 2020. PMID: 32582680 Free PMC article.
-
Rib Presence, Anterior Rib Cage Integrity, and Segmental Length Affect the Stability of the Human Thoracic Spine: An in vitro Study.Front Bioeng Biotechnol. 2020 Feb 7;8:46. doi: 10.3389/fbioe.2020.00046. eCollection 2020. Front Bioeng Biotechnol. 2020. PMID: 32117927 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials
Miscellaneous