Intravascular imaging comparison of two metallic limus-eluting stents abluminally coated with biodegradable polymers: IVUS and OCT results of the DESTINY trial

Int J Cardiovasc Imaging. 2017 Feb;33(2):161-168. doi: 10.1007/s10554-016-0992-z. Epub 2016 Oct 6.

Abstract

We sought to compare, by means of IVUS and OCT imaging, the performance of a novel sirolimus-eluting drug-eluting stent (DES) with biodegradable polymer (Inspiron™) to the Biomatrix™ DES. From the DESTINY trial, a total of 70 randomized patients (2:1) were enrolled in the IVUS substudy (Inspiron™, n = 46; Biomatrix™: n = 20) while 25 patients were evaluated with OCT (Inspiron™, n = 19; Biomatrix™: n = 06) at 9-month follow-up. The main endpoints were % of neointimal tissue obstruction (IVUS) and neointimal stut coverage (OCT) at 9 months. Patients treated with both DES had very little NIH formation at 9 months either by IVUS (% of NIH obstruction of 4.9 ± 4.1 % with Inspiron™ vs. 2.7 ± 2.9 % with Biomatrix™, p = 0.03) or by OCT (neointimal thickness of 144.2 ± 72.5 µm Inspiron™ vs. 115.0 ± 53.9 µm with Biomatrix™, p = 0.45). Regarding OCT strut-level assessment, again both devices showed excellent 9-month performance, with high rates of strut coverage (99.49 ± 1.01 % with Inspiron™ vs. 97.62 ± 2.21 % with Biomatrix™, p < 0.001) and very rare malapposition (0.29 ± 1.06 % with Inspiron™ vs. 0.53 ± 0.82 % with Biomatrix™, p = 0.44). Patients with any uncovered struts were more frequently identified in the Biomatrix™ group (9.78 ± 7.13 vs. 2.29 ± 3.91 %, p < 0.001). In the present study, midterm IVUS and OCT evaluations showed that both new generation DES with biodegradable polymer were effective in terms of suppressing excessive neointimal response, with very high rates of apposed and covered struts, suggesting a consistent and benign healing pattern.

Keywords: Biodegradable polymer; Biolimus A9; Drug-eluting stents; Sirolimus.

Publication types

  • Clinical Trial
  • Comparative Study

MeSH terms

  • Absorbable Implants*
  • Cardiovascular Agents / administration & dosage*
  • Cardiovascular Agents / adverse effects
  • Coronary Artery Disease / diagnostic imaging*
  • Coronary Artery Disease / therapy*
  • Coronary Vessels / diagnostic imaging*
  • Coronary Vessels / drug effects*
  • Drug-Eluting Stents*
  • Humans
  • Metals*
  • Neointima
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Polymers / chemistry*
  • Predictive Value of Tests
  • Prosthesis Design
  • Sirolimus / administration & dosage
  • Sirolimus / adverse effects
  • Sirolimus / analogs & derivatives*
  • Time Factors
  • Tomography, Optical Coherence*
  • Treatment Outcome
  • Ultrasonography, Interventional*

Substances

  • Cardiovascular Agents
  • Metals
  • Polymers
  • umirolimus
  • Sirolimus