Lead extraction outcomes in patients with congenital heart disease

Europace. 2017 Mar 1;19(3):441-446. doi: 10.1093/europace/euw049.

Abstract

Aims: Patients with congenital heart disease (CHD) are at increased risk for intracardiac device malfunction and infection that may necessitate extraction; however, the risk of extraction is poorly understood. This study addresses the safety of extraction in patients with structural heart disease and previous cardiac surgery.

Methods and results: This retrospective study included 40 CHD and 80 matched control patients, who underwent transvenous lead extractions between 2001 and 2014. Only leads >12 months were included. There were 77 leads in CHD patients and 146 in controls. The mean age was 38 ± 16 years in CHD patients. Ninety per cent of CHD patients had ≥1 cardiac surgeries when compared with 21% of controls (P < 0.001). The number of abandoned leads was significantly different (17 vs. 3, P < 0.001). Lead age was similar with an average duration of 83 ± 87 months in CHD patients and 62 ± 65 months in controls (P = 0.24). There was no significant difference in extraction techniques. Manual traction was successful in 40% of CHD patients and 47% of controls, and advanced techniques were used in 60 and 53% of CHD patients and controls, respectively. Complete extraction was achieved in 94% of the patients in both groups. There was no significant difference in complications.

Conclusion: Lead extraction can be safely performed in patients with CHD. Despite anatomic abnormalities and longer implantation times, the difficulty of lead extraction in patients with CHD is comparable with controls.

Keywords: Cardiac implantable electronic device; Congenital heart disease; Laser lead extraction; Lead extraction.

Publication types

  • Multicenter Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Arrhythmias, Cardiac / diagnosis
  • Arrhythmias, Cardiac / etiology
  • Arrhythmias, Cardiac / therapy*
  • Cardiac Pacing, Artificial* / adverse effects
  • Cardiac Surgical Procedures* / adverse effects
  • Child
  • Defibrillators, Implantable
  • Device Removal* / adverse effects
  • Electric Countershock / adverse effects
  • Electric Countershock / instrumentation*
  • Female
  • Heart Defects, Congenital / complications
  • Heart Defects, Congenital / diagnosis
  • Heart Defects, Congenital / surgery*
  • Heart Failure / diagnosis
  • Heart Failure / etiology
  • Heart Failure / therapy*
  • Humans
  • Male
  • Middle Aged
  • Minnesota
  • Oregon
  • Pacemaker, Artificial*
  • Postoperative Complications / etiology
  • Prosthesis Failure*
  • Retrospective Studies
  • Risk Assessment
  • Risk Factors
  • Time Factors
  • Treatment Outcome
  • Ventricular Dysfunction / diagnosis
  • Ventricular Dysfunction / etiology
  • Ventricular Dysfunction / therapy*
  • Young Adult