Voxel-Based Sensitivity of Flat-Panel CT for the Detection of Intracranial Hemorrhage: Comparison to Multi-Detector CT

PLoS One. 2016 Nov 2;11(11):e0165794. doi: 10.1371/journal.pone.0165794. eCollection 2016.

Abstract

Objectives: Flat-panel CT (FPCT) allows cross-sectional parenchymal, vascular and perfusion imaging within the angiography suite, which could greatly facilitate acute stroke management. We hypothesized that FPCT offers equal diagnostic accuracy compared to multi-detector CT (MDCT) as a primary tool to exclude intracranial hemorrhage.

Methods: 22 patients with intracranial hematomas who had both MDCT and FPCT performed within 24 hours were retrospectively identified. Patients with visible change in hematoma size or configuration were excluded. Two raters independently segmented hemorrhagic lesions. Data sets and corresponding binary lesion maps were co-registered to compare hematoma volume. Diagnostic accuracy of FPCT to detect hemorrhage was calculated from voxel-wise analysis of lesion overlap compared to reference MDCT.

Results: Mean hematoma size was similar between MDCT (16.2±8.9 ml) and FPCT (16.1±8.6 ml), with near perfect correlation of hematoma sizes between modalities (ρ = 0.95, p<0.001). Sensitivity and specificity of FPCT to detect hemorrhagic voxels was 61.6% and 99.8% for intraventricular hematomas and 67.7% and 99.5% for all other intracranial hematomas.

Conclusions: In this small sample containing predominantly cases with subarachnoid hemorrhage, FPCT based assessment of hemorrhagic volume in brain yields acceptable accuracy compared to reference MDCT, albeit with a limited sensitivity on a voxel level. Further assessment and improvement of FPCT is necessary before it can be applied as a primary imaging modality to exclude intracranial hemorrhage in acute stroke patients.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Aged
  • Female
  • Humans
  • Intracranial Hemorrhages / diagnostic imaging*
  • Male
  • Middle Aged
  • Radiographic Image Enhancement / methods*
  • Retrospective Studies
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / instrumentation*
  • Tomography, X-Ray Computed / methods

Grants and funding