Reliability of pulse waveform separation analysis: effects of posture and fasting
- PMID: 27861248
- DOI: 10.1097/HJH.0000000000001182
Reliability of pulse waveform separation analysis: effects of posture and fasting
Abstract
Objective: Oscillometric pulse wave analysis devices enable, with relative simplicity and objectivity, the measurement of central hemodynamic parameters. The important parameters are central blood pressures and indices of arterial wave reflection, including wave separation analysis (backward pressure component Pb and reflection magnitude). This study sought to determine whether the measurement precision (between-day reliability) of Pb and reflection magnitude: exceeds the criterion for acceptable reliability; and is affected by posture (supine, seated) and fasting state.
Methods: Twenty healthy adults (50% female, 27.9 years, 24.2 kg/m) were tested on six different mornings: 3 days fasted, 3 days nonfasted condition. On each occasion, participants were tested in supine and seated postures. Oscillometric pressure waveforms were recorded on the left upper arm.
Results: The criterion intra-class correlation coefficient value of 0.75 was exceeded for Pb (0.76) and reflection magnitude (0.77) when participants were assessed under the combined supine-fasted condition. The intra-class correlation coefficient was lowest for Pb in seated-nonfasted condition (0.57), and lowest for reflection magnitude in the seated-fasted condition (0.56). For Pb, the smallest detectible change that must be exceeded in order for a significant change to occur in an individual was 2.5 mmHg, and for reflection magnitude, the smallest detectable change was 8.5%.
Conclusion: Assessments of Pb and reflection magnitude are as follows: exceed the criterion for acceptable reliability; and are most reliable when participants are fasted in a supine position. The demonstrated reliability suggests sufficient precision to detect clinically meaningful changes in reflection magnitude and Pb.
Similar articles
-
Reliability of oscillometric central blood pressure and wave reflection readings: effects of posture and fasting.J Hypertens. 2015 Aug;33(8):1588-93. doi: 10.1097/HJH.0000000000000604. J Hypertens. 2015. PMID: 26136065
-
Reliability of oscillometric central blood pressure and central systolic loading in individuals over 50 years: Effects of posture and fasting.Atherosclerosis. 2018 Feb;269:79-85. doi: 10.1016/j.atherosclerosis.2017.12.030. Epub 2017 Dec 27. Atherosclerosis. 2018. PMID: 29339276
-
Reliability of oscillometric central hemodynamic responses to an orthostatic challenge.Atherosclerosis. 2015 Aug;241(2):761-5. doi: 10.1016/j.atherosclerosis.2015.06.041. Epub 2015 Jun 27. Atherosclerosis. 2015. PMID: 26142686
-
Clinical measurement of arterial stiffness obtained from noninvasive pressure waveforms.Am J Hypertens. 2005 Jan;18(1 Pt 2):3S-10S. doi: 10.1016/j.amjhyper.2004.10.009. Am J Hypertens. 2005. PMID: 15683725 Review.
-
Pulse Waveform Analysis: Is It Ready for Prime Time?Curr Hypertens Rep. 2017 Aug 11;19(9):73. doi: 10.1007/s11906-017-0769-3. Curr Hypertens Rep. 2017. PMID: 28801878 Review.
Cited by
-
Metformin improves skeletal muscle microvascular insulin resistance in metabolic syndrome.Am J Physiol Endocrinol Metab. 2022 Feb 1;322(2):E173-E180. doi: 10.1152/ajpendo.00287.2021. Epub 2021 Dec 27. Am J Physiol Endocrinol Metab. 2022. PMID: 34957859 Free PMC article.
-
A Primer on Repeated Sitting Exposure and the Cardiovascular System: Considerations for Study Design, Analysis, Interpretation, and Translation.Front Cardiovasc Med. 2021 Aug 17;8:716938. doi: 10.3389/fcvm.2021.716938. eCollection 2021. Front Cardiovasc Med. 2021. PMID: 34485414 Free PMC article. Review.
-
The impact of upper-limb position on estimated central blood pressure waveforms.J Hum Hypertens. 2019 Jun;33(6):444-453. doi: 10.1038/s41371-019-0179-x. Epub 2019 Feb 25. J Hum Hypertens. 2019. PMID: 30804459
-
Influence of Exercise Mode on Post-exercise Arterial Stiffness and Pressure Wave Measures in Healthy Adult Males.Front Physiol. 2018 Oct 17;9:1468. doi: 10.3389/fphys.2018.01468. eCollection 2018. Front Physiol. 2018. PMID: 30459633 Free PMC article.
-
Oscillometric central blood pressure and central systolic loading in stroke patients: Short-term reproducibility and effects of posture and fasting state.PLoS One. 2018 Nov 1;13(11):e0206329. doi: 10.1371/journal.pone.0206329. eCollection 2018. PLoS One. 2018. PMID: 30383781 Free PMC article. Clinical Trial.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical