Customized mandibular reconstruction plates improve mechanical performance in a mandibular reconstruction model

Comput Methods Biomech Biomed Engin. 2017 Mar;20(4):426-435. doi: 10.1080/10255842.2016.1240788. Epub 2016 Nov 25.

Abstract

The purpose of this paper was to analyze the biomechanical performance of customized mandibular reconstruction plates with optimized strength. The best locations for increasing bar widths were determined with a sensitivity analysis. Standard and customized plates were mounted on mandible models and mechanically tested. Maximum stress in the plate could be reduced from 573 to 393 MPa (-31%) by increasing bar widths. The median fatigue limit was significantly greater (p < 0.001) for customized plates (650 ± 27 N) than for standard plates (475 ± 27 N). Increasing bar widths at case-specific locations was an effective strategy for increasing plate fatigue performance.

Keywords: Mandibular reconstruction; fatigue testing; finite element analysis; implant; sensitivity analysis.

MeSH terms

  • Biomechanical Phenomena
  • Bone Plates*
  • Finite Element Analysis
  • Humans
  • Mandible / surgery
  • Mandibular Reconstruction / methods*
  • Stress, Mechanical