Reduced radiation dose and improved image quality using a mini mobile digital imaging system in a neonatal intensive care unit

Clin Imaging. 2017 Mar-Apr:42:165-171. doi: 10.1016/j.clinimag.2016.12.004. Epub 2016 Dec 6.

Abstract

This study was aimed to assess the radiation dose and image quality of a mini-mobile digital imaging (mini-DI) system for neonatal chest radiography and compared to conventional digital radiography (DR). A total of 64 neonates were examined and anatomical landmarks were assessed. The entrance surface dose of mini DI and conventional DR was 26.64±0.15 μGy and 49.11±1.46 μGy, respectively (p<0.001). The mean SNR values for mini-DI and DR were 233.2±5.1 and 31.6±1.2, and 10% MTF values were 131 and 161μm. A newly developed mini-DI is capable of preserving the diagnostic information with dose reduction in neonates under intensive care.

Keywords: Chest X-ray imaging; Conventional mobile digital radiography (DR); Mini-mobile digital imaging system (mini-DI); Premature neonate.

MeSH terms

  • Humans
  • Infant, Newborn
  • Intensive Care Units, Neonatal*
  • Phantoms, Imaging
  • Radiation Dosage
  • Radiographic Image Enhancement / methods*
  • Radiography, Thoracic / methods*
  • Thorax / diagnostic imaging*