A Swiss cheese error detection method for real-time EPID-based quality assurance and error prevention
Abstract
Purpose: To develop a robust and efficient process that detects relevant dose errors (dose errors of ≥5%) in external beam radiation therapy and directly indicates the origin of the error. The process is illustrated in the context of electronic portal imaging device (EPID)-based angle-resolved volumetric-modulated arc therapy (VMAT) quality assurance (QA), particularly as would be implemented in a real-time monitoring program.
Methods: A Swiss cheese error detection (SCED) method was created as a paradigm for a cine EPID-based during-treatment QA. For VMAT, the method compares a treatment plan-based reference set of EPID images with images acquired over each 2° gantry angle interval. The process utilizes a sequence of independent consecutively executed error detection tests: an aperture check that verifies in-field radiation delivery and ensures no out-of-field radiation; output normalization checks at two different stages; global image alignment check to examine if rotation, scaling, and translation are within tolerances; pixel intensity check containing the standard gamma evaluation (3%, 3 mm) and pixel intensity deviation checks including and excluding high dose gradient regions. Tolerances for each check were determined. To test the SCED method, 12 different types of errors were selected to modify the original plan. A series of angle-resolved predicted EPID images were artificially generated for each test case, resulting in a sequence of precalculated frames for each modified treatment plan. The SCED method was applied multiple times for each test case to assess the ability to detect introduced plan variations. To compare the performance of the SCED process with that of a standard gamma analysis, both error detection methods were applied to the generated test cases with realistic noise variations.
Results: Averaged over ten test runs, 95.1% of all plan variations that resulted in relevant patient dose errors were detected within 2° and 100% within 14° (<4% of patient dose delivery). Including cases that led to slightly modified but clinically equivalent plans, 89.1% were detected by the SCED method within 2°. Based on the type of check that detected the error, determination of error sources was achieved. With noise ranging from no random noise to four times the established noise value, the averaged relevant dose error detection rate of the SCED method was between 94.0% and 95.8% and that of gamma between 82.8% and 89.8%.
Conclusions: An EPID-frame-based error detection process for VMAT deliveries was successfully designed and tested via simulations. The SCED method was inspected for robustness with realistic noise variations, demonstrating that it has the potential to detect a large majority of relevant dose errors. Compared to a typical (3%, 3 mm) gamma analysis, the SCED method produced a higher detection rate for all introduced dose errors, identified errors in an earlier stage, displayed a higher robustness to noise variations, and indicated the error source.
Keywords: EPID; VMAT; error detection; real-time quality assurance; treatment verification.
© 2017 American Association of Physicists in Medicine.
Similar articles
-
Commissioning and quality assurance for VMAT delivery systems: An efficient time-resolved system using real-time EPID imaging.Med Phys. 2017 Aug;44(8):3909-3922. doi: 10.1002/mp.12387. Epub 2017 Jul 10. Med Phys. 2017. PMID: 28564208
-
An error detection method for real-time EPID-based treatment delivery quality assurance.Med Phys. 2021 Feb;48(2):569-578. doi: 10.1002/mp.14633. Epub 2020 Dec 30. Med Phys. 2021. PMID: 33314247
-
A novel method for sub-arc VMAT dose delivery verification based on portal dosimetry with an EPID.Med Phys. 2017 Nov;44(11):5556-5562. doi: 10.1002/mp.12518. Epub 2017 Sep 18. Med Phys. 2017. PMID: 28815696
-
Is it necessary to repeat quality control procedures for head and neck patients?Radiother Oncol. 1991 Jul;21(3):201-10. doi: 10.1016/0167-8140(91)90038-i. Radiother Oncol. 1991. PMID: 1924856 Review.
-
In vivo dosimetry in external beam photon radiotherapy: Requirements and future directions for research, development, and clinical practice.Phys Imaging Radiat Oncol. 2020 Aug 29;15:108-116. doi: 10.1016/j.phro.2020.08.003. eCollection 2020 Jul. Phys Imaging Radiat Oncol. 2020. PMID: 33458335 Free PMC article. Review.
Cited by
-
A recurrent neural network for rapid detection of delivery errors during real-time portal dosimetry.Phys Imaging Radiat Oncol. 2022 Apr 20;22:36-43. doi: 10.1016/j.phro.2022.03.004. eCollection 2022 Apr. Phys Imaging Radiat Oncol. 2022. PMID: 35493850 Free PMC article.
-
A method for in vivo treatment verification of IMRT and VMAT based on electronic portal imaging device.Radiat Oncol. 2021 Dec 4;16(1):232. doi: 10.1186/s13014-021-01953-9. Radiat Oncol. 2021. PMID: 34863229 Free PMC article.
-
Virtual Patient-Specific Quality Assurance of IMRT Using UNet++: Classification, Gamma Passing Rates Prediction, and Dose Difference Prediction.Front Oncol. 2021 Jul 20;11:700343. doi: 10.3389/fonc.2021.700343. eCollection 2021. Front Oncol. 2021. PMID: 34354949 Free PMC article.
-
Optimisation of a composite difference metric for prompt error detection in real-time portal dosimetry of simulated volumetric modulated arc therapy.Br J Radiol. 2021 Apr 1;94(1120):20201014. doi: 10.1259/bjr.20201014. Epub 2021 Mar 18. Br J Radiol. 2021. PMID: 33733813 Free PMC article.
-
Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.Phys Imaging Radiat Oncol. 2018 Mar 19;5:76-84. doi: 10.1016/j.phro.2018.03.002. eCollection 2018 Jan. Phys Imaging Radiat Oncol. 2018. PMID: 33458373 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources