Quality in dual-energy X-ray absorptiometry scans
Abstract
Dual-energy X-ray absorptiometry (DXA) is the gold standard for measuring bone mineral density (BMD), making the diagnosis of osteoporosis, and for monitoring changes in BMD over time. DXA data are also used in the determination of fracture risk. Procedural steps in DXA scanning can be broken down into scan acquisition, analysis, interpretation, and reporting. Careful attention to quality control pertaining to these procedural steps should theoretically be beneficial in patient management. Inattention to procedural steps and errors that may occur at each step has the possibility of providing information that would inform inappropriate clinical decisions, generating unnecessary healthcare expenses and ultimately causing avoidable harm to patients. This article reviews errors in DXA scanning that affect trueness and precision related to the machine, the patient, and the technologist and reviews articles which document problems with DXA quality in clinical and research settings. An understanding of DXA errors is critical for DXA quality; programs such as certification of DXA technologists and interpreters help in assuring quality bone densitometry. As DXA errors are common, pay for performance requiring DXA technologists and interpreters to be certified and follow quality indicators is indicated.
Keywords: Artifacts; Dual-energy X-ray absorptiometry; Errors; Quality; Scan acquisition; Scan analysis.
Copyright © 2017 Elsevier Inc. All rights reserved.
Similar articles
-
Best Practices for Dual-Energy X-ray Absorptiometry Measurement and Reporting: International Society for Clinical Densitometry Guidance.J Clin Densitom. 2016 Apr-Jun;19(2):127-40. doi: 10.1016/j.jocd.2016.03.003. Epub 2016 Mar 22. J Clin Densitom. 2016. PMID: 27020004
-
Common errors in dual-energy X-ray absorptiometry scans in imaging centers in Ecuador.Arch Osteoporos. 2020 Jan 2;15(1):6. doi: 10.1007/s11657-019-0673-3. Arch Osteoporos. 2020. PMID: 31897808
-
DXA Errors Are Common and Reduced by Use of a Reporting Template.J Clin Densitom. 2019 Jan-Mar;22(1):115-124. doi: 10.1016/j.jocd.2018.07.014. Epub 2018 Sep 13. J Clin Densitom. 2019. PMID: 30327243
-
Dual-energy X-ray absorptiometry in clinical practice: application and interpretation of scans beyond the numbers.Clin Imaging. 2002 Jan-Feb;26(1):43-9. doi: 10.1016/s0899-7071(01)00356-4. Clin Imaging. 2002. PMID: 11814753 Review.
-
Fundamentals and pitfalls of bone densitometry using dual-energy X-ray absorptiometry (DXA).Osteoporos Int. 2004 Nov;15(11):847-54. doi: 10.1007/s00198-004-1681-7. Epub 2004 Aug 21. Osteoporos Int. 2004. PMID: 15322740 Review.
Cited by
-
Bone Loss and Fractures in Post-Menopausal Women Living with HIV: A Narrative Review.Pathogens. 2024 Sep 19;13(9):811. doi: 10.3390/pathogens13090811. Pathogens. 2024. PMID: 39339002 Free PMC article. Review.
-
PET/CT for the Opportunistic Screening of Osteoporosis and Fractures in Cancer Patients.Curr Osteoporos Rep. 2024 Dec;22(6):553-560. doi: 10.1007/s11914-024-00887-x. Epub 2024 Sep 14. Curr Osteoporos Rep. 2024. PMID: 39276167 Review.
-
Asymmetric Osteopenia in Adolescent Idiopathic Scoliosis Based on Hounsfield Unit of Computed Tomography.Int J Gen Med. 2024 Sep 7;17:3945-3953. doi: 10.2147/IJGM.S478933. eCollection 2024. Int J Gen Med. 2024. PMID: 39263592 Free PMC article.
-
A literature review of biomarkers used for diagnosis of relative energy deficiency in sport.Front Sports Act Living. 2024 Jul 12;6:1375740. doi: 10.3389/fspor.2024.1375740. eCollection 2024. Front Sports Act Living. 2024. PMID: 39070233 Free PMC article.
-
Medial Calcar Density Measured via Opportunistic Computed Tomography Is Well Represented by the Dorr C Classification.Cureus. 2024 Jun 15;16(6):e62428. doi: 10.7759/cureus.62428. eCollection 2024 Jun. Cureus. 2024. PMID: 39011184 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials