Configurations of the Re-scan Confocal Microscope (RCM) for biomedical applications

J Microsc. 2017 May;266(2):166-177. doi: 10.1111/jmi.12526. Epub 2017 Mar 3.

Abstract

The new high-sensitive and high-resolution technique, Re-scan Confocal Microscopy (RCM), is based on a standard confocal microscope extended with a re-scan detection unit. The re-scan unit includes a pair of re-scanning mirrors that project the emission light onto a camera in a scanning manner. The signal-to-noise ratio of Re-scan Confocal Microscopy is improved by a factor of 4 compared to standard confocal microscopy and the lateral resolution of Re-scan Confocal Microscopy is 170 nm (compared to 240 nm for diffraction limited resolution, 488 nm excitation, 1.49 NA). Apart from improved sensitivity and resolution, the optical setup of Re-scan Confocal Microscopy is flexible in its configuration in terms of control of the mirrors, lasers and filters. Because of this flexibility, the Re-scan Confocal Microscopy can be configured to address specific biological applications. In this paper, we explore a number of possible configurations of Re-scan Confocal Microscopy for specific biomedical applications such as multicolour, FRET, ratio-metric (e.g. pH and intracellular Ca2+ measurements) and FRAP imaging.

Keywords: High-sensitive microscopy; live cell imaging; multicolour microscopy; super-resolution microscopy.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Cell Line
  • Cytological Techniques / instrumentation*
  • Cytological Techniques / methods*
  • Humans
  • Microscopy, Confocal / instrumentation*
  • Microscopy, Confocal / methods*