Patient dose in interventional radiology: a multicentre study of the most frequent procedures in France

Eur Radiol. 2017 Oct;27(10):4281-4290. doi: 10.1007/s00330-017-4780-5. Epub 2017 Mar 13.

Abstract

Objectives: A national retrospective survey on patient doses was performed by the French Society of Medical physicists to assess reference levels (RLs) in interventional radiology as required by the European Directive 2013/59/Euratom.

Methods: Fifteen interventional procedures in neuroradiology, vascular radiology and osteoarticular procedures were analysed. Kerma area product (KAP), fluoroscopy time (FT), reference air kerma and number of images were recorded for 10 to 30 patients per procedure. RLs were calculated as the 3rd quartiles of the distributions.

Results: Results on 4600 procedures from 36 departments confirmed the large variability in patient dose for the same procedure. RLs were proposed for the four dosimetric estimators and the 15 procedures. RLs in terms of KAP and FT were 90 Gm.cm2 and 11 mins for cerebral angiography, 35 Gy.cm2 and 16 mins for biliary drainage, 75 Gy.cm2 and 6 mins for lower limbs arteriography and 70 Gy.cm2 and 11 mins for vertebroplasty. For these four procedures, RLs were defined according to the complexity of the procedure. For all the procedures, the results were lower than most of those already published.

Conclusions: This study reports RLs in interventional radiology based on a national survey. Continual evolution of practices and technologies requires regular updates of RLs.

Key points: • Delivered dose in interventional radiology depends on procedure, practice and patient. • National RLs are proposed for 15 interventional procedures. • Reference levels (RLs) are useful to benchmark practices and optimize protocols. • RLs are proposed for kerma area product, air kerma, fluoroscopy time and number of images. • RLs should be adapted to the procedure complexity and updated regularly.

Keywords: Interventional radiology; Patient dose; Patient dosimetry; Radiation protection; Reference levels.