Recognition-mediated particle detection under microfluidic flow with waveguide-coupled 2D photonic crystals: towards integrated photonic virus detectors

Lab Chip. 2017 May 2;17(9):1570-1577. doi: 10.1039/c7lc00221a.

Abstract

Label-free biodetection schemes compatible with standard CMOS fabrication methods constitute an important goal, as these are enabling tools for the mass production of high-sensitivity biosensors. Two-dimensional slab photonic crystal (2D slab-PhC) sensors have been posited as ultrahigh-sensitivity detection components, but to date recognition-mediated detection of viruses or simulants under flow has not been demonstrated. We report the design and optimization of a new W1 waveguide-coupled 2D slab-PhC sensor, with a geometry well suited to virus detection. Proof of concept experiments with fluorescent latex particles verified that the sensor could respond to infiltration of a single particle, both in air and under an aqueous cover layer. Subsequent experiments with antibody-functionalized sensors and virus simulants confirmed the ability of the device to detect virus-sized particles under flow via a recognition-mediated process. This work sets the stage for incorporation of 2D slab-PhC sensors into fully integrated photonic sensor systems.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Antibodies, Viral / metabolism
  • Biosensing Techniques / instrumentation*
  • Humans
  • Microfluidic Analytical Techniques / instrumentation*
  • Microfluidic Analytical Techniques / methods*
  • Optics and Photonics / instrumentation*
  • Viruses / isolation & purification*
  • Viruses / metabolism

Substances

  • Antibodies, Viral