A Novel 141Ce-Based Flood Field Phantom: Assessment of Suitability for Daily Uniformity Testing in a Clinical Nuclear Medicine Department

J Nucl Med Technol. 2017 Sep;45(3):225-229. doi: 10.2967/jnmt.116.189340. Epub 2017 Apr 13.

Abstract

Daily quality control testing of a γ-camera is of the utmost importance in assessing whether the camera is suitable for clinical use. The aim of our study was to assess the suitability of a fillable 141Ce-based flood field phantom developed in-house for daily quality control testing of γ-cameras. Methods: Daily uniformity testing was performed for 113 d using the fillable 141Ce phantom and a commercially available sheet-type 57Co phantom, and the results were compared. Results: The average integral uniformity obtained by the 141Ce and 57Co phantoms was 3.24% and 2.72%, respectively, for detector 1 and 3.31% and 2.78%, respectively, for detector 2. Conclusion: The 141Ce phantom we developed is a suitable alternative to the commercially available 57Co phantom.

Keywords: 141Ce flood field uniformity; 57Co flood field uniformity; flood field phantom; gamma camera; surrogate radionuclide.