Plain pelvic radiographs are commonly used for a variety of pediatric orthopedic disorders. Lead shielding is typically placed over the gonads to minimize radiation exposure to these sensitive tissues. However, misplaced shielding can sabotage efforts to protect patients from excessive radiation exposure either by not covering radiosensitive tissues or by obscuring anatomic areas of interest, prompting repeat radiographic examinations. The goal of this study was to determine the incidence of misplaced shielding for pelvic radiographs obtained for pediatric orthopedic evaluation. Children 8 to 16 years old who had an anteroposterior or frog lateral pelvic radiograph between 2008 and 2014 were included. A total of 3400 patients met the inclusion criteria, and 84 boys and 84 girls were randomly selected for review. For both boys and girls, the percentage of incorrectly positioned or missing shields was calculated. Chi-square testing was used to compare the frequency of missing or incorrectly placed shields between sexes and age groups. Pelvic shields were misplaced in 49% of anteroposterior and 63% of frog lateral radiographs. Shielding was misplaced more frequently for girls than for boys on frog lateral radiographs (76% vs 51%; P<.05). Pelvic bony landmarks were often obscured by pelvic shielding, with a frequency of 7% to 43%, depending on the specific landmark. The femoral head and acetabulum were obscured by shielding in up to 2% of all images. The findings suggest that accepted pelvic shielding protocols are ineffective. Consideration should be given to alternative protocols or abandonment of this practice. [Orthopedics. 2017; 40(4):e623-e627.].
Copyright 2017, SLACK Incorporated.