, 12 (8), 1333-1343

Grasper Having Tactile Sensing Function Using Acoustic Reflection for Laparoscopic Surgery

Affiliations

Grasper Having Tactile Sensing Function Using Acoustic Reflection for Laparoscopic Surgery

Hiep Hoang Ly et al. Int J Comput Assist Radiol Surg.

Abstract

Purpose: In current minimally invasive surgery techniques, the tactile information available to the surgeon is limited. Improving tactile sensation could enhance the operability of surgical instruments. Considering surgical applications, requirements such as having electrical safety, a simple structure, and sterilization capability should be considered. The current study sought to develop a grasper that can measure grasping force at the tip, based on a previously proposed tactile sensing method using acoustic reflection. This method can satisfy the requirements for surgical applications because it has no electrical element within the part that is inserted into the patient's body.

Methods: We integrated our acoustic tactile sensing method into a conventional grasping forceps instrument. We designed the instrument so that acoustic cavities within a grasping arm and a fork sleeve were connected by a small cavity in a pivoting joint. In this design, when the angle between the two grasping arms changes during grasping, the total length and local curvature of the acoustic cavity remain unchanged. Thus, the grasping force can be measured regardless of the orientation of the grasping arm.

Results: We developed a prototype sensorized grasper based on our proposed design. Fundamental tests revealed that sensor output increased with increasing contact force applied to the grasping arm, and the angle of the grasping arm did not significantly affect the sensor output. Moreover, the results of a grasping test, in which objects with different softness characteristics were held by the grasper, revealed that the grasping force could be appropriately adjusted to handle different objects on the basis of sensor output.

Conclusions: Experimental results demonstrated that the prototype grasper can measure grasping force, enabling safe and stable grasping.

Keywords: Acoustic reflection; Grasper; Minimally invasive surgery; Tactile sensor.

Similar articles

See all similar articles

Cited by 2 PubMed Central articles

  • Recent Advances in Tactile Sensing Technology
    M Park et al. Micromachines (Basel) 9 (7). PMID 30424254. - Review
    Research on tactile sensing technology has been actively conducted in recent years to pave the way for the next generation of highly intelligent devices. Sophisticated ta …
  • Robotic Surgery - Who Is The Boss?
    PJ Schuler. Laryngorhinootologie 97 (S 01), S231-S278. PMID 29905359. - Review
    In the head and neck region, great potential is seen in robot-assisted surgery (RAS). Mainly in cancer surgery, the use of robotic systems seems to be of interest. Until …

References

    1. Int J Comput Assist Radiol Surg. 2015 Feb;10(2):183-93 - PubMed
    1. J Biomech Eng. 2008 Apr;130(2):021020 - PubMed
    1. Int J Med Robot. 2010 Jun;6(2):221-30 - PubMed
    1. Conf Proc IEEE Eng Med Biol Soc. 2014;2014:6517-20 - PubMed
    1. Conf Proc IEEE Eng Med Biol Soc. 2012;2012:944-7 - PubMed

LinkOut - more resources

Feedback