Objective: The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure monitor YuWell YE690A for blood pressure measurement according to the International Protocol of the European Society of Hypertension revision 2010.
Participants and methods: Systolic and diastolic blood pressures were measured sequentially in 33 adult Chinese (12 women, 44.2 years of mean age) using a mercury sphygmomanometer (two observers) and the YE690A device (one supervisor). A total of 99 pairs of comparisons were obtained from 33 participants for judgments in two parts with three grading phases.
Results: All the blood pressure requirements were fulfilled. The YuWell YE690A device achieved the targets in part 1 of the validation study. The number of absolute differences between device and observers within 5, 10, and 15 mmHg was 79/99, 96/99, and 97/99, respectively, for systolic blood pressure and 72/99, 95/99, and 98/99, respectively, for diastolic blood pressure. The device also fulfilled the criteria in part 2 of the validation study. Thirty-one and 25 participants for systolic and diastolic blood pressure, respectively, had at least two of the three device-observer differences within 5 mmHg (required ≥24). No participant for systolic and two participants for diastolic blood pressure had all the three device-observer comparisons greater than 5 mmHg.
Conclusion: The YuWell blood pressure monitor YE690A has passed the requirements of the International Protocol revision 2010 and hence can be recommended for blood pressure measurement in adults.