Validation of the YuWell YE690A upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010

Blood Press Monit. 2017 Oct;22(5):295-297. doi: 10.1097/MBP.0000000000000278.

Abstract

Objective: The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure monitor YuWell YE690A for blood pressure measurement according to the International Protocol of the European Society of Hypertension revision 2010.

Participants and methods: Systolic and diastolic blood pressures were measured sequentially in 33 adult Chinese (12 women, 44.2 years of mean age) using a mercury sphygmomanometer (two observers) and the YE690A device (one supervisor). A total of 99 pairs of comparisons were obtained from 33 participants for judgments in two parts with three grading phases.

Results: All the blood pressure requirements were fulfilled. The YuWell YE690A device achieved the targets in part 1 of the validation study. The number of absolute differences between device and observers within 5, 10, and 15 mmHg was 79/99, 96/99, and 97/99, respectively, for systolic blood pressure and 72/99, 95/99, and 98/99, respectively, for diastolic blood pressure. The device also fulfilled the criteria in part 2 of the validation study. Thirty-one and 25 participants for systolic and diastolic blood pressure, respectively, had at least two of the three device-observer differences within 5 mmHg (required ≥24). No participant for systolic and two participants for diastolic blood pressure had all the three device-observer comparisons greater than 5 mmHg.

Conclusion: The YuWell blood pressure monitor YE690A has passed the requirements of the International Protocol revision 2010 and hence can be recommended for blood pressure measurement in adults.

Publication types

  • Validation Study

MeSH terms

  • Adult
  • Blood Pressure Monitoring, Ambulatory / instrumentation*
  • Blood Pressure*
  • Female
  • Humans
  • Male
  • Observer Variation
  • Sphygmomanometers / standards*