Purpose: Implement and evaluate a 3D MRI method to measure temperature changes with high spatial and temporal resolution and large field of view.
Methods: A multiecho pseudo-golden angle stack-of-stars (SOS) sequence with k-space weighted image contrast (KWIC) reconstruction was implemented to simultaneously measure multiple quantities, including temperature, initial signal magnitude M(0), transverse relaxation time ( T2*), and water/fat images. Respiration artifacts were corrected using self-navigation. KWIC artifacts were removed using a multi-baseline library. The phases of the multiple echo images were combined to improve proton resonance frequency precision. Temperature precision was tested through in vivo breast imaging (N = 5 healthy volunteers) using both coronal and sagittal orientations and with focused ultrasound (FUS) heating in a pork phantom using a breast specific MR-guided FUS system.
Results: Temperature measurement precision was significantly improved after echo combination when compared with the no echo combination case (spatial average of the standard deviation through time of 0.3-1.0 and 0.7-1.9°C, respectively). Temperature measurement accuracy during heating was comparable to a 3D seg-EPI sequence. M(0) and T2* values showed temperature dependence during heating in pork adipose tissue.
Conclusion: A self-navigated 3D multiecho SOS sequence with dynamic KWIC reconstruction is a promising thermometry method that provides multiple temperature sensitive quantitative values. Magn Reson Med 79:1407-1419, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
Keywords: KWIC; MRgFUS; stack of stars; thermometry.