Neopuff T-piece resuscitator mask ventilation: Does mask leak vary with different peak inspiratory pressures in a manikin model?

J Paediatr Child Health. 2017 Aug;53(8):761-765. doi: 10.1111/jpc.13609. Epub 2017 Jul 4.

Abstract

Aim: The aim of this study was to compare mask leak with three different peak inspiratory pressure (PIP) settings during T-piece resuscitator (TPR; Neopuff) mask ventilation on a neonatal manikin model.

Methods: Participants were neonatal unit staff members. They were instructed to provide mask ventilation with a TPR with three PIP settings (20, 30, 40 cm H2 O) chosen in a random order. Each episode was for 2 min with 2-min rest period. Flow rate and positive end-expiratory pressure (PEEP) were kept constant. Airway pressure, inspiratory and expiratory tidal volumes, mask leak, respiratory rate and inspiratory time were recorded. Repeated measures analysis of variance was used for statistical analysis.

Results: A total of 12 749 inflations delivered by 40 participants were analysed. There were no statistically significant differences (P > 0.05) in the mask leak with the three PIP settings. No statistically significant differences were seen in respiratory rate and inspiratory time with the three PIP settings. There was a significant rise in PEEP as the PIP increased. Failure to achieve the desired PIP was observed especially at the higher settings.

Conclusions: In a neonatal manikin model, the mask leak does not vary as a function of the PIP when the flow rate is constant. With a fixed rate and inspiratory time, there seems to be a rise in PEEP with increasing PIP.

Keywords: mask leak; mask ventilation; neonatology.

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis / methods
  • Intensive Care Units, Neonatal
  • Manikins*
  • Masks / standards*
  • New South Wales
  • Positive-Pressure Respiration*
  • Tidal Volume