Evaluation of Accuracy of Six Blood Glucose Monitoring Systems and Modeling of Possibly Related Insulin Dosing Errors

Diabetes Technol Ther. 2017 Oct;19(10):580-588. doi: 10.1089/dia.2016.0408. Epub 2017 Jul 13.

Abstract

Objective: Self-monitoring of blood glucose (BG) is an essential part of diabetes therapy. Accurate and reliable results from BG monitoring systems (BGMS) are important especially when they are used to calculate insulin doses. This study aimed at assessing system accuracy of BGMS and possibly related insulin dosing errors.

Research design and methods: System accuracy of six different BGMS (Accu-Chek® Aviva Nano, Accu-Chek Mobile, Accu-Chek Performa Nano, CONTOUR® NEXT LINK 2.4, FreeStyle Lite, OneTouch® Verio® IQ) was assessed in comparison to a glucose oxidase and a hexokinase method. Study procedures and analysis were based on ISO 15197:2013/EN ISO 15197:2015, clause 6.3. In addition, insulin dosing error was modeled.

Results: In the comparison against the glucose oxidase method, five out of six BGMS fulfilled ISO 15197:2013 accuracy criteria. Up to 14.3%/4.3%/0.3% of modeled doses resulted in errors exceeding ±0.5/±1.0/±1.5 U and missing the modeled target by 20 mg/dL/40 mg/dL/60 mg/dL, respectively. Compared against the hexokinase method, five out of six BGMS fulfilled ISO 15197:2013 accuracy criteria. Up to 25.0%/10.5%/3.2% of modeled doses resulted in errors exceeding ±0.5/±1.0/±1.5 U, respectively.

Conclusions: Differences in system accuracy were found, even among BGMS that fulfilled the minimum system accuracy criteria of ISO 15197:2013. In the error model, considerable insulin dosing errors resulted for some of the investigated systems. Diabetes patients on insulin therapy should be able to rely on their BGMS' readings; therefore, they require highly accurate BGMS, in particular, when making therapeutic decisions.

Keywords: Blood glucose monitoring; ISO 15197; Insulin dosing error; Self-monitoring of blood glucose; System accuracy.

Publication types

  • Evaluation Study

MeSH terms

  • Blood Glucose / analysis*
  • Blood Glucose Self-Monitoring / instrumentation*
  • Blood Glucose Self-Monitoring / methods
  • Diabetes Mellitus / blood*
  • Diabetes Mellitus / drug therapy
  • Female
  • Humans
  • Hypoglycemic Agents / administration & dosage*
  • Hypoglycemic Agents / therapeutic use
  • Insulin / administration & dosage*
  • Insulin / therapeutic use
  • Male
  • Reproducibility of Results

Substances

  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin