New Mydriasis-Free Electroretinogram Recorded with Skin Electrodes in Healthy Subjects
- PMID: 28713831
- PMCID: PMC5496106
- DOI: 10.1155/2017/8539747
New Mydriasis-Free Electroretinogram Recorded with Skin Electrodes in Healthy Subjects
Abstract
Purpose: To evaluate the reproducibility and consistency of the new mydriasis-free electroretinogram (ERG) with a skin electrode (RETeval) device, to determine the normative values of parameters, and to clarify the usefulness of pupil records to colored-light stimulus.
Methods: A total of 100 eyes of 50 healthy subjects (mean age, 21.4 years) were enrolled. The diagnostic parameters obtained by the RETeval device were examined under the following conditions. The reproducibility was determined with the coefficient of variation (CV). The consistency was examined by intraclass correlation coefficients (ICCs). The mean value and the normal range were analyzed with a 95% confidence interval as the normative values of parameters. The correlation of parameters to pupil records (area ratio, constriction ratio) and flicker ERG was also examined in the diabetic retinopathy assessment protocol.
Results: From the CV for each of the two measurements, the amplitude has a low reproducibility compared with the implicit time. Generally good consistency was obtained with both ERG parameters (ICCs = 0.48-0.92). Moderate correlations were found for the white-, red-, and blue-light stimulus in the area ratio and the constriction ratio, respectively (r = 0.44-0.62; P = 0.010-<0.0001). No correlation was observed between pupil and flicker parameters (r = 0.06-0.34; P = 0.646-0.051).
Conclusions: The RETeval device was suggested as a possible screening device to detect the visual afferent diseases by evaluating in combination with the ERG recording and the colored-light pupil response.
Figures
Similar articles
-
Evaluation of light- and dark-adapted ERGs using a mydriasis-free, portable system: clinical classifications and normative data.Doc Ophthalmol. 2018 Dec;137(3):169-181. doi: 10.1007/s10633-018-9660-z. Epub 2018 Oct 24. Doc Ophthalmol. 2018. PMID: 30357588
-
Role of a mydriasis-free, full-field flicker ERG device in the detection of diabetic retinopathy.Doc Ophthalmol. 2018 Dec;137(3):131-141. doi: 10.1007/s10633-018-9656-8. Epub 2018 Oct 17. Doc Ophthalmol. 2018. PMID: 30334119
-
Screening for diabetic retinopathy using new mydriasis-free, full-field flicker ERG recording device.Sci Rep. 2016 Nov 8;6:36591. doi: 10.1038/srep36591. Sci Rep. 2016. PMID: 27824158 Free PMC article. Clinical Trial.
-
Factors Affecting Mydriasis-Free Flicker ERGs Recorded With Real-Time Correction for Retinal Illuminance: Study of 150 Young Healthy Subjects.Invest Ophthalmol Vis Sci. 2017 Oct 1;58(12):5280-5286. doi: 10.1167/iovs.17-22587. Invest Ophthalmol Vis Sci. 2017. PMID: 29049731
-
Electrodes and the recording of the human electroretinogram (ERG).Int J Psychophysiol. 1994 May;16(2-3):131-6. doi: 10.1016/0167-8760(89)90039-1. Int J Psychophysiol. 1994. PMID: 8089031 Review.
Cited by
-
Paediatric norms for photopic electroretinogram testing based on a large cohort of Chinese preschool children.BMJ Open Ophthalmol. 2024 Feb 21;9(1):e001393. doi: 10.1136/bmjophth-2023-001393. BMJ Open Ophthalmol. 2024. PMID: 38388004 Free PMC article.
-
Full-field electroretinogram recorded with skin electrodes in 6- to 12-year-old children.Doc Ophthalmol. 2023 Dec;147(3):179-188. doi: 10.1007/s10633-023-09944-9. Epub 2023 Aug 2. Doc Ophthalmol. 2023. PMID: 37530953 Free PMC article.
-
Investigating the Structural and Functional Changes in the Optic Nerve in Patients with Early Glaucoma Using the Optical Coherence Tomography (OCT) and RETeval System.Sensors (Basel). 2023 May 5;23(9):4504. doi: 10.3390/s23094504. Sensors (Basel). 2023. PMID: 37177707 Free PMC article.
-
Comparing the RETeval® portable ERG device with more traditional tabletop ERG systems in normal subjects and selected retinopathies.Doc Ophthalmol. 2023 Apr;146(2):137-150. doi: 10.1007/s10633-022-09903-w. Epub 2022 Oct 23. Doc Ophthalmol. 2023. PMID: 36273363 Free PMC article.
-
Novel Machine-Learning Based Framework Using Electroretinography Data for the Detection of Early-Stage Glaucoma.Front Neurosci. 2022 May 4;16:869137. doi: 10.3389/fnins.2022.869137. eCollection 2022. Front Neurosci. 2022. PMID: 35600610 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical