Initial assessment of a novel delivery system (NuDEL™ ®) for the covered Cheatham-Platinum stent

Cardiol Young. 2017 Oct;27(8):1465-1469. doi: 10.1017/S1047951117000488.

Abstract

Aims We sought to evaluate the first-in-man use of a new system for implantation of covered stents in patients with complex structural and CHD. Methods and results Retrospective data were collected of the first 13 NuDEL™ delivery systems used in patients. The NuDEL™ comprises a covered Cheatham-Platinum stent mounted on a balloon-in-balloon and pre-loaded in a long delivery sheath. Data were collected from three centres in the United Kingdom and Ireland. A total of 13 covered stents were delivered via 12 NuDEL™ delivery systems in 12 patients. Among them, six patients had coarctation of the aorta, five patients had right ventricular outflow tract stenosis, and one patient had severe stenosis of a Mustard systemic venous baffle. There were no complications, and all the stents were deployed in the desired position with satisfactory haemodynamic results.

Conclusions: The development of a bespoke system of a pre-mounted, pre-loaded covered stent may negate some of the technical challenges that complicate large-calibre stent deployment. Our preliminary results suggest that the NuDEL™ system is a safe and effective means of covered stent deployment in challenging anatomy.

Keywords: Covered stent; coarctation; percutaneous pulmonary valve; right ventricular outflow tract stent.

Publication types

  • Multicenter Study

MeSH terms

  • Adolescent
  • Adult
  • Aortic Coarctation / diagnosis
  • Aortic Coarctation / surgery*
  • Aortography
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Child
  • Coated Materials, Biocompatible*
  • Equipment Design
  • Female
  • Humans
  • Male
  • Platinum / pharmacology
  • Prosthesis Design
  • Retrospective Studies
  • Stents*
  • Treatment Outcome
  • Young Adult

Substances

  • Coated Materials, Biocompatible
  • Platinum