Commissioning of a PTW 34070 large-area plane-parallel ionization chamber for small field megavoltage photon dosimetry

J Appl Clin Med Phys. 2017 Nov;18(6):206-217. doi: 10.1002/acm2.12185. Epub 2017 Oct 4.

Abstract

Purpose: This study investigates a large-area plane-parallel ionization chamber (LAC) for measurements of dose-area product in water (DAPw ) in megavoltage (MV) photon fields.

Methods: Uniformity of electrode separation of the LAC (PTW34070 Bragg Peak Chamber, sensitive volume diameter: 8.16 cm) was measured using high-resolution microCT. Signal dependence on angle α of beam incidence for square 6 MV fields of side length s = 20 cm and 1 cm was measured in air. Polarity and recombination effects were characterized in 6, 10, and 18 MV photons fields. To assess the lateral setup tolerance, scanned LAC profiles of a 1 × 1 cm2 field were acquired. A 6 MV calibration coefficient, ND,w,LAC , was determined in a field collimated by a 5 cm diameter stereotactic cone with known DAPw . Additional calibrations in 10 × 10 cm2 fields at 6, 10, and 18 MV were performed.

Results: Electrode separation is uniform and agrees with specifications. Volume-averaging leads to a signal increase proportional to ~1/cos(α) in small fields. Correction factors for polarity and recombination range between 0.9986 to 0.9996 and 1.0007 to 1.0024, respectively. Off-axis displacement by up to 0.5 cm did not change the measured signal in a 1 × 1 cm2 field. ND,w,LAC was 163.7 mGy cm-2 nC-1 and differs by +3.0% from the coefficient derived in the 10 × 10 cm2 6 MV field. Response in 10 and 18 MV fields increased by 1.0% and 2.7% compared to 6 MV.

Conclusions: The LAC requires only small correction factors for DAPw measurements and shows little energy dependence. Lateral setup errors of 0.5 cm are tolerated in 1 × 1 cm2 fields, but beam incidence must be kept as close to normal as possible. Calibration in 10 × 10 fields is not recommended because of the LAC's over-response. The accuracy of relative point-dose measurements in the field's periphery is an important limiting factor for the accuracy of DAPw measurements.

Keywords: dose-area product; large-area chamber; linear accelerators; quality assurance; small field dosimetry.